38
RovaP
series
parallel
evaporator
can
realize
use
of
different
volumes
of
sample
bottles
by
replacing
the
upper
cover
plate
and
sample
adapter.
◆
Replace cover plate
Turn off power of equipment, making cover plate in a closed state, remove countersunk
head screws on the left and right sides of cover plate, pull out the pins, take off cover plate
and replace it.
Note: The upper surface of cover plate is made of glass, and must be handled with care to
avoid bump damage!
The surface of the upper cover is coated with Teflon coating to avoid bumping and
scratching the surface!
◆
Replace cover plate adapter
Remove the fixing screws and replace sample bottle adapter.