24
◆ 【Method】
Establishment ,preservation and loading of sample concentration methods
◆ 【Save】
The currently set parameters such as temperature, time, vacuum, and
oscillation speed can be saved to method;
◆【Setting】
Language and vacuum degree can be chosen;
Preheat mode can be turned on or off;
User management is possible;
Check running records and equipment information;
Heating temperature of cover plate can be set;
Water injection control is possible.
◆ 【Lock】
The vacuum degree can be locked in current state, when you do not know
target vacuum degree, vacuum degree target value can be set to a smaller
value, and then start equipment to concentrate sample. When there have
liquid to entering the collection bottle, quickly click the [Lock] button,
equipment will automatically keep vacuum constant in the current state;
When equipment runs in step vacuum mode, if you press [Lock] button or
use adjustment knob to change the vacuum degree target value,
equipment will exit gradient vacuum degree mode and run at the
currently set vacuum target.
◆ 【Release 】
After clicking button, sample continues to oscillate, but vacuum pump
stops working, equipment will release vacuum, and cover plate will