9.2.1.1 Method setting interface
◆
【Name】
Please enter the method name to be created;
◆
【Creator】
Automatically created according to the user's name;
◆
【Temp.】
According to requirements of analysis method, set heating temperature
of water bath, temperature setting range: room tempe5 °C to 85 °C;
◆
【Speed】
Set sample shaking speed. Faster shaking speed can shorten time required for
sample concentration, generally recommended value is 150 to 200rpm. When
the load-bearing capacity of the worktable is more than or equal to 500kg, a
higher oscillation speed can be set, otherwise it will cause the worktable to
vibrate;
◆
【Storage area code】
The storage area code is automatically displayed;
◆
【Step control 】
The lad which can prevent vacuum from rising too fast ,avoiding sample
explosive boiling.
It supports up to 5-stage vacuum control, allowing sample to be concentrated
in stages according to the set vacuum and duration; when multi-stage