11. RS232 PC SERIAL INTERFACE

The instrument has RS232 PC serial interface via a 3.5
mm terminal ( 3-12, Fig. 1 ).

The data output is a 16 digit stream which can be
utilized  for user's specific application.

A RS232 lead with the following connection will be
required to link the instrument with the PC serial port.

Meter

PC

(9W 'D" Connector)

Center Pin............................................

Pin 4

(3.5 mm jack plug)
 Ground/shield...........................................

Pin 2

 2.2 K
resistor

Pin 5

The 16 digits data stream will be displayed in the
following  format :

D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0

Each digit indicates the following status :

D0

End Word = 0D

D1 & D8

Display reading,  D8 = MSD , D1 = LSD
For example : 

If the display reading is 1234, then D8 to
D1 is : 00001234

D9

Decimal Point(DP), position from right to the left
0 = No DP, 1= 1 DP, 2 = 2 DP, 3 = 3 DP

D10

Polarity
0 = Positive

1 = Negative

D11 & D12

Annunciator for Display
mT = E3

G = B5

D13

When send the upper display data = 1

D14

4

D15

Start Word

20

Summary of Contents for MG-3003SD

Page 1: ...nt Although this meter is a complex and delicate instrument its durable structure will allow many years of use if proper operating techniques are developed Please read the following instructions caref...

Page 2: ...e 0 second 11 5 4 Check time information 11 5 5 SD Card Data structure 12 6 Saving data from the SD card to the computer 13 7 ADVANCED SETTING 15 7 1 Set clock time Year Month Date Hour Minute Second...

Page 3: ...D memory card Datalogger it Built in Clock and Calendar real time data recorder sampling time set from 1 second to 3600 seconds Manual datalogger is available set the sampling time to 0 during execute...

Page 4: ...5 AC 5 rdg 20 digit at 23 5 Frequency AC measurement Response 50 Hz 60 Hz Sensor Hall sensor Datalogger Auto 1 sec to 3600 sec Sampling Time Sampling time can set to 1 second Setting range but memory...

Page 5: ...Power Current Normal operation w o SD card save data and LCD Backlight is OFF Approx DC 13 mA When SD card save the data but and LCD Backlight is OFF Approx DC 35 mA If LCD backlight on the power con...

Page 6: ...ter Input Socket 3 4 REC Button 3 15 Battery Compartment Cover 3 5 Range Button Button 3 16 Battery Cover Screws 3 6 Function Button Button 3 17 Stand 3 7 Set Button Time Button 3 18 Tripod Fix Nut 3...

Page 7: ...ing the FUNCTION Button 3 6 Fig 1 until function DC change to AC or function AC change to DC 3 magnetic field measuremen range select by pressing and holding the Range Button 3 5 Fig 1 until Range 1 c...

Page 8: ...surement Display will show AC indicator Display will not show N S indicator 5 Contact and close the testing material to the probe head refer to Fig 3 the Display will show the magnetic field value alo...

Page 9: ...will release the data hold function 4 3 Data Record Max Min reading 1 The data record function records the maximum and minimum readings Press the REC Button 3 4 Fig 1 once to start the Data Record fu...

Page 10: ...xecute the RECORD function it can not change from the North pole to South pole or from the South pole to North pole during the measurement It can make the measurement under the same pole S pole or N p...

Page 11: ...n formatted by other meter or by other installation such as camera Reformat the memory card with your meter If the SD memory card exist the trouble during format by the meter use the computer to refor...

Page 12: ...4 page 17 b Pause the datalogger During execute the Datalogger function if press the LOG Button 3 8 Fig 1 once will pause the Datalogger function stop to save the measuring data into the memory circui...

Page 13: ...3 5 Fig 1 or Button 3 6 Fig 1 to set the measuring position 1 to 99 for example room 1 to room 99 to identify the measurement location the Top Display will show P x x 1 to 99 b Finish the Datalogger D...

Page 14: ...w file name MGA01001 XLS After exist the Datalogger then execute again the data will save to the MGA01001 XLS until Data column reach to 30 000 columns then will generate a new file for example MGA010...

Page 15: ...D card into the SD card adapter then connect the SD card adapter into the computer 3 Power ON the computer and run the EXCEL software Down load the saving data file for example the file name MGA01001...

Page 16: ...EXCEL graphic screen for example 14...

Page 17: ...r SD F SD memory card Format UNIT Select the magnetic field unit to mT or G Remark During execute the Advanced Setting function if press SET Button 3 7 Fig 1 1 5 seconds will exit the Advanced Setting...

Page 18: ...5 Fig 1 or Button 3 6 Fig 1 to adjust the value 0 1 2 5 10 30 60 120 300 600 1800 3600 seconds 2 After the Sampling value is selected press the Enter Button 3 8 Fig 1 will save the setting function wi...

Page 19: ...ting The numerical data structure of SD card is default used the as the decimal for example 20 6 1000 53 But in certain countries Europe is used the as the decimal point for example 20 6 1000 53 Under...

Page 20: ...R to confirm again if make sure to do the SD memory card format then press Enter Button once will format the SD memory clear all the existing data that already saving into the SD card 7 7 Select the m...

Page 21: ...ator appears before the instrument become inaccurate 2 Loose the screws of the Battery Cover 3 16 Fig 1 and take away the Battery Cover 3 15 Fig 1 from the instrument and remove the battery 3 Replace...

Page 22: ...ck plug Ground shield Pin 2 2 2 K resistor Pin 5 The 16 digits data stream will be displayed in the following format D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0 Each digit indicates the foll...

Page 23: ...8 Data bits Stop bit 1 Stop bit 12 PATENT The meter SD card structure already get patent or patent pending in following countries Germany Nr 20 2008 016 337 4 JAPAN 3151214 TAIWAN M 358970 M 359043 C...

Reviews: