Water Purification System                                                                                                      Instruction Manual 

  11 / 11 

.

 Motor working but no water 

RO membrane is blocked or pump’s pressure lost. For the former, change the RO membrane; For the later, 

check whether expiring of pump is caused by long time overload or blocking of the cartridge. Repair motor. 

 

.Motor works ever and again 

The motor works ever and again. It is the reason that low pressure or blocking of LV.1cartridge.  Please 

install a booster pump or replace LV.1cartridge.   

 

.The pressure tank is full of water, but pure water can not flow out 

A.  Air in the pressure tank leaks. Please fill it with 7psi air and check whether air leaks. If air can not be 

put into tank, change the pressure tank.. 

B.  Backset active carbon is blocked. Please replace it with a new one. 

 

. Power is off, but RO water does not stop flowing. 

A.  Check the four-side water circuit breaker, and replace it if it goes wrong. 

B.  Pressure  of  inlet  water  is  higher  then  the  set  pressure  of  high  pressure  switch.  Please  regulate  the 

pressure of inlet water no more then 2kg/cm2.   

 

. The machine can not stop running when pressure tank is full of water or it runs at once after one stop. 

A.  High pressure switch or liquid level controller goes wrong. Repair it if it works after repairing, otherwise, 

replace it with a new one. 

B.  Non-return valve leaks pressure so that it can not close. Replace it please. 

 

11.

 

 

Warranty & Repair Regulation 

The products customers buy form our company enjoy repair service since the day of purchase. In on year 

form  the  purchasing  day,  our  company  is  obliged  to  replace  components  free  of  charge  due  to 

non-human-behavior factors except: 

(1).  Damage caused by maloperation or use in abnormal situations. 

(2).  Disassembly any part of the machine or human-behavior damage. 

(3).  Not repaired by our serviceman. 

 

 

Specification can be changed without any prior notice for development 

 

 

Summary of Contents for Master-S UVF

Page 1: ...ater Purification System Instruction Manual Model Master S UVF The instruction manual should be read before installation Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC...

Page 2: ...day Model Master S UVF Quality of Ultrapure water Resistivity 18 2 M cm TOC 3ppb Endotoxin 0 001EU ml Bacteria 1 CFU ml Particle 0 1 m 1 ml Heavy metal ion 0 1ppb Quality of RO water TDS ppm mg l TDS...

Page 3: ...uality easy put adapter Proper usage is as follows Connect the tube press the oval cap of the interface strongly then insert the tube to the bottom of adapter Take off the tube press the oval cap of t...

Page 4: ...of machine 3 Connect to RO drain Use 1 4 PE tube with a suitable length One side connects with the interface with black label marked To drain at the back of machine and the other side is directed to g...

Page 5: ...uatic products feeding inlet water for ultra pure water machine inlet water for sterilizer t h chamber Terminal filter Tap water Tap water adapter 2 Tap water adapter 1 3 direction links Plastic ball...

Page 6: ...Ultrapure water higher quality water than RO water Standby When RO water and Ultrapure water is not for use the system will be in standby state The system still produce RO water to store in the water...

Page 7: ...xecute corresponding function UP key press open or close Ultrapure water valve Remarks The initial code is 1000 After pressing MENU key press please in code will appear Inputting the code 1000 through...

Page 8: ...until the state of inlet water is ok And system will produce pure water again C When system is under working conditions it will warn for no inlet water or low pressure of inlet water The sound and pi...

Page 9: ...r limit alarm parameter of TDS1 Remark TDS1 hysteresis of inlet water conductivity Max Min upper and lower alarm test only work when power is on When current test parameter TDS1 s Max the current test...

Page 10: ...system s initial code is 1000 Under the menu interface picture 15 press MENU key press to shift the cursor to choose Code item then press OK key press to enter Code setting picture 19 Press key press...

Page 11: ...trapure polishing resin cartridge 4 About 4000 liter pure water UF 5000D 5000 doulton uf cartridge 1 TF 0 45 0 1 m S 0 45 0 1 m terminal filter 1 Remarks Worse inlet source water quality or big dosage...

Page 12: ...side water circuit breaker and replace it if it goes wrong B Pressure of inlet water is higher then the set pressure of high pressure switch Please regulate the pressure of inlet water no more then 2k...

Reviews: