Water Purification System                                                                                                      Instruction Manual 

  9 / 11 

resistivity < 

The “Min”, system will warn to replace the Ultrapure polishing resin cartridge. When 

current test parameter of resistivity < 

The “Min”, system will stop warning. 

EC: electrode parameter setting (Parameter setting is OK in factory. Additional parameter setting is 

unnecessary.) 

(3). 

Parts’ life setting / parts’ runtime clear 

Under the menu interface(picture 15), press “MENU” key-press to shift the cursor to choose “Parts” 
item, then press “OK” key-press to enter ”Parts” setting(picture 18). 

Pre-treat: pre-

treat’s lift setting. (range: 0-9999 hours) 

UV lamp: UV lamp’s life setting. (range: 0-9999 hours 

 

 

 

picture 18

 

Pre-treat runtime: clr - Pre-treat runtime clear.   

UV lamp runtime: clr - UV lamp runtime clear. 
(press “MENU” key-press to shift the cursor to choose corresponding “clr” icon, then press “OK” 

key-press to clear runtime. And system will restart timing) 

(4).  Code setting 

All system’s initial code is “1000” 
Under the menu interface(picture 15), press “MENU” key-press to shift the cursor to choose “Code” 
item, then press “OK” key-press to enter ”Code” setting(picture 19). Press “      ” key-press to choose 
modified place, and press “      ” key-press to amend the parameter. After amending code, system will 

use new code for entering menu interface. 

 

 

 

picture 19 

(5).  RO water, Ultrapure water timing dispense setting 

Under the menu interface(picture 15), press “MENU” key-press to shift the cursor to choose 
“dispense” item, then press “OK” key-press to enter ”dispense” setting(picture 20). 2 kind dispense 

mode is optional-

“Manual” and “Auto”. When system enters “Auto” dispense mode, timing setting is 

viable. 

UP out: Ultrapure 

water timing dispense setting (only for “Auto” dispense mode) 

RO out: RO water timing dispense setting (only for “Auto” dispense mode) 

 

 

 

picture 20 

 

8.

 

Water Quality Test 

The system has 3 monitor sensors of water quality measuring.   

 

The first, TDS1, 

monitors feed water’s quality (tap water, inlet water).   

Measure unit: TDS(total dissolved solid, ppm) 

 

The second, TDS2,

 

monitors reverse osmosis water’s quality(RO water).   

Summary of Contents for Master-S UVF

Page 1: ...ater Purification System Instruction Manual Model Master S UVF The instruction manual should be read before installation Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC...

Page 2: ...day Model Master S UVF Quality of Ultrapure water Resistivity 18 2 M cm TOC 3ppb Endotoxin 0 001EU ml Bacteria 1 CFU ml Particle 0 1 m 1 ml Heavy metal ion 0 1ppb Quality of RO water TDS ppm mg l TDS...

Page 3: ...uality easy put adapter Proper usage is as follows Connect the tube press the oval cap of the interface strongly then insert the tube to the bottom of adapter Take off the tube press the oval cap of t...

Page 4: ...of machine 3 Connect to RO drain Use 1 4 PE tube with a suitable length One side connects with the interface with black label marked To drain at the back of machine and the other side is directed to g...

Page 5: ...uatic products feeding inlet water for ultra pure water machine inlet water for sterilizer t h chamber Terminal filter Tap water Tap water adapter 2 Tap water adapter 1 3 direction links Plastic ball...

Page 6: ...Ultrapure water higher quality water than RO water Standby When RO water and Ultrapure water is not for use the system will be in standby state The system still produce RO water to store in the water...

Page 7: ...xecute corresponding function UP key press open or close Ultrapure water valve Remarks The initial code is 1000 After pressing MENU key press please in code will appear Inputting the code 1000 through...

Page 8: ...until the state of inlet water is ok And system will produce pure water again C When system is under working conditions it will warn for no inlet water or low pressure of inlet water The sound and pi...

Page 9: ...r limit alarm parameter of TDS1 Remark TDS1 hysteresis of inlet water conductivity Max Min upper and lower alarm test only work when power is on When current test parameter TDS1 s Max the current test...

Page 10: ...system s initial code is 1000 Under the menu interface picture 15 press MENU key press to shift the cursor to choose Code item then press OK key press to enter Code setting picture 19 Press key press...

Page 11: ...trapure polishing resin cartridge 4 About 4000 liter pure water UF 5000D 5000 doulton uf cartridge 1 TF 0 45 0 1 m S 0 45 0 1 m terminal filter 1 Remarks Worse inlet source water quality or big dosage...

Page 12: ...side water circuit breaker and replace it if it goes wrong B Pressure of inlet water is higher then the set pressure of high pressure switch Please regulate the pressure of inlet water no more then 2k...

Reviews: