Water Purification System                                                                                                      Instruction Manual 

  5 / 11 

6.

 

Usage Guide 

All data have been set in the factory. The machine will operate smoothly without any data-setting. All work state 

will display in the LCD. If there is abnormal state, the system will alarm automatically. If data modification is 

necessary, specific step is in 

“Microcomputer Controller”. 

 

 

The power switch and 

“on-off duty” work model switch are at the back of the shell.   

Specific picture is as follows: 

 

                                             

 

 

 
“on-off duty” work model switch is as follows.

 

(Remarks:”On duty” work mode is recommended)

 

 

 

 

 

 

                                     

 

Starting up: 

Turn on the tap water valve, insert the power line into the power source and turn on the power switch, then 

the system begins to produce pure water.   

 

 

Getting corresponding pure water: 

Press “RO” or ”UP” key-press, which are on the panel, to get corresponding RO water or Ultrapure 

water(higher quality water than RO water).   

 

 

Standby: 

When RO water and Ultrapure water is not for use, the system will be in standby state. The system still 

produce RO water to store in the water tank

(optional)

. Until tank is full, the system will automatically stop. 

The system will begin to produce pure water again when any pure water is used. 

 

 

Shutdown: 

Turn off the tap water valve and turn off the power switch. Then it is ok. 

 

 

Flushing UF membrane 

Method of flushing UF membrane:   

Turn on the UF drain valve, and press 

”UP” key-press, which are on the panel, to turn on Ultrapure water’s 

valve 

Attention: The 

“UF drain valve” is shut on normal condition, except for flushing UF membrane. 

Frequency of flushing UF membrane:   

At least one times every week, and at least listing 30 seconds every times. 

Attention: Make sure that the water source and power source is not connected when the system is 

not in the use state for long time(for example, off duty).   

On duty model 

Stand by 

Off duty model 

Turn on the power 
 
Turn off the power 

Summary of Contents for Master-S UVF

Page 1: ...ater Purification System Instruction Manual Model Master S UVF The instruction manual should be read before installation Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC...

Page 2: ...day Model Master S UVF Quality of Ultrapure water Resistivity 18 2 M cm TOC 3ppb Endotoxin 0 001EU ml Bacteria 1 CFU ml Particle 0 1 m 1 ml Heavy metal ion 0 1ppb Quality of RO water TDS ppm mg l TDS...

Page 3: ...uality easy put adapter Proper usage is as follows Connect the tube press the oval cap of the interface strongly then insert the tube to the bottom of adapter Take off the tube press the oval cap of t...

Page 4: ...of machine 3 Connect to RO drain Use 1 4 PE tube with a suitable length One side connects with the interface with black label marked To drain at the back of machine and the other side is directed to g...

Page 5: ...uatic products feeding inlet water for ultra pure water machine inlet water for sterilizer t h chamber Terminal filter Tap water Tap water adapter 2 Tap water adapter 1 3 direction links Plastic ball...

Page 6: ...Ultrapure water higher quality water than RO water Standby When RO water and Ultrapure water is not for use the system will be in standby state The system still produce RO water to store in the water...

Page 7: ...xecute corresponding function UP key press open or close Ultrapure water valve Remarks The initial code is 1000 After pressing MENU key press please in code will appear Inputting the code 1000 through...

Page 8: ...until the state of inlet water is ok And system will produce pure water again C When system is under working conditions it will warn for no inlet water or low pressure of inlet water The sound and pi...

Page 9: ...r limit alarm parameter of TDS1 Remark TDS1 hysteresis of inlet water conductivity Max Min upper and lower alarm test only work when power is on When current test parameter TDS1 s Max the current test...

Page 10: ...system s initial code is 1000 Under the menu interface picture 15 press MENU key press to shift the cursor to choose Code item then press OK key press to enter Code setting picture 19 Press key press...

Page 11: ...trapure polishing resin cartridge 4 About 4000 liter pure water UF 5000D 5000 doulton uf cartridge 1 TF 0 45 0 1 m S 0 45 0 1 m terminal filter 1 Remarks Worse inlet source water quality or big dosage...

Page 12: ...side water circuit breaker and replace it if it goes wrong B Pressure of inlet water is higher then the set pressure of high pressure switch Please regulate the pressure of inlet water no more then 2k...

Reviews: