Product Front View

Product Side View

11

12

6. Operational Instruction

● 

     heat dissipation outlet and direct sunshine should be avoided.

● 

Adjust the instrument footings to level the instrument and make sure the instrument

     can operate smoothly.

● 

Please make sure the instrument is placed at lease 50 cm away from the wall and

     other objects.

● 

Please use a dedicated power outlet with a capacity no less than 1.3 kW and AC 220 V.

● 

It is not allowed to use gasoline, banana oil, and any other volatile chemicals to contact

     with the outside coating of the instrument.

● 

No experiment samples are allowed in the control panel.

● 

No excessive force should be used when opening and closing the incubator door to

    prevent the instrument from being damaged.

● 

Please reduce the times of opening the incubator door to help maintain constant

     temperature and save CO

2

.

● 

After the instrument cooling function is used for 10 days continuously, a heating

     dehumidification process is recommended. Recommendation: heat the instrument to

     45 °C and last for half an hour.

● 

Inspect the fastening screws on the flask clamp regularly to prevent it from falling off.

● 

Please clean the instrument regularly.

● 

Please check if the v-belt at the bottom of the instrument is not loosened after using the

     instrument for 6-12 months. Press the middle part of the v-belt inward about 10 mm.

     Please check the abrasion condition of the v-belt after using the instrument for 2 years

     and change the same type of v-belt (Z1000) if necessary.

● 

Please disassembly the side plate at the right and back of the instrument to clean out

      dust on the condenser and fan to ensure good heat dissipation and high cooling

      efficiency.

● 

Dehumidification is required if the humidity of the instrument is too high.

● 

Ensure good ventilation of the room.

 Please make sure the air circulation is sufficient when using CO

2

. The CO

2

    concentration will increase in a small confined room and do harm to your health. Also,

    please avoid direct inhalation of the CO

2

 gas when opening the the incubator door.

The instrument should be palced in a clean and dry environment. Any heat source or

7. Structure

Temperature and
speed control panel

 CO

2

 control panel

Handle

Ventilation hole

Adjustable footings

Infrared lamp switch

Power switch

Fluorescent lamp
switch

Filter

 CO2 switch

Power inlet

Summary of Contents for IncuShaker LOM-7450-CO2

Page 1: ...Stackable Incubator Shaker User s Manual Please read this user s manual carefully before using the product LOM 7450 7450 CO2 7450 L...

Page 2: ...choosing IncuShaker Series Incubator Shaker Your trust and support in MRC LTD are greatly appreciated MRC Incubator Shaker Series is model designed for shaking and heat samples according to the resear...

Page 3: ...e instruments are running with the maximum load it is recommended to set the rotation speed no more than 250 rpm 11 Please make sure the vent of the instrument is kept away from other vents or heat so...

Page 4: ...rview 2 2 Features Large view window LCD dot matrix display User friendly interface Multi segment programming Extra long timing up to 999 hours Soft start and smooth acceleration Open door cut off swi...

Page 5: ...Range Platform Size Platform Number Maximum Capacity Standard Configuration Power Power Supply Overall Size Net Weight CO2 Concentration Rang CO2 Concentration Precision CO2 Detecting sensor Chamber H...

Page 6: ...X H c Overal size when stacking three units W X D X H 124 78 67cm 124 132 67cm 124 78 200cm Please leave adequate space around the instrument to help the cooling system of the instrument to dissipate...

Page 7: ...shown in figure 2 It may cause the pipe connect the incubator and the inlet port to fall off and the leak of the CO2 gas if the secondary pressure is too high High CO2 concentration will do harm to yo...

Page 8: ...tening screws on the flask clamp regularly to prevent it from falling off Please clean the instrument regularly Please check if the v belt at the bottom of the instrument is not loosened after using t...

Page 9: ...to stop the instrument when it s in Simple Operation or Timing Program mode 9 Settings Turn on the power switch at the right side of the instrument after the instrument is properly placed Thanks for...

Page 10: ...s after finishing running the program Return to Timing Program page When Timing Program Operation shows on the display window press the button to choose whether to continue running Stoppig Y N Press t...

Page 11: ...ing Main Interface Press the button to choose a function and enter the button to enter the corresponding operation page a Basic settings Press the button to choose the parameter that needs to be modif...

Page 12: ...e current parameters and return to the temperature control parameter settings C 2 2 Temperature Calibration The value outside of the bracket is the target temperature value the value inside the bracke...

Page 13: ...he calibrated value are not saved If press the button in the calibration mode the instrument will exit the CO2 calibration and the target value is not saved and reset to factory settings The instrumen...

Page 14: ...for more than 5 minutes The alarm indicator light will flash and the buzz will sound The electromagnetic gas valve will be closed automatically Press any key to cancel the alarm status and the instru...

Page 15: ...t connected Power outlet out of power Power switch off Change to a new fuse with the same specification Position at the powr outlet Position inside the electrical cabinet Remove interference source of...

Page 16: ...For Model LOM 7450 CO2 only For Model LOM 7450 CO2 only For Model LOM 7450 CO2 only For Model LOM 7450 CO2 only For Model LOM 7450 CO2 only Optional Accessories Name Specifications Name Specifications...