6. Operational Instruction
●
heat dissipation outlet and direct sunshine should be avoided.
●
Adjust the instrument footings to level the instrument and make sure the instrument
can operate smoothly.
●
Please make sure the instrument is placed at lease 50 cm away from the wall and
other objects.
●
Please use a dedicated power outlet with a capacity no less than 1.3 kW and AC 220 V.
●
It is not allowed to use gasoline, banana oil, and any other volatile chemicals to contact
with the outside coating of the instrument.
●
No experiment samples are allowed in the control panel.
●
No excessive force should be used when opening and closing the incubator door to
prevent the instrument from being damaged.
●
Please reduce the times of opening the incubator door to help maintain constant
temperature and save CO
2
.
●
After the instrument cooling function is used for 10 days continuously, a heating
dehumidification process is recommended. Recommendation: heat the instrument to
45 °C and last for half an hour.
●
Inspect the fastening screws on the flask clamp regularly to prevent it from falling off.
●
Please clean the instrument regularly.
●
Please check if the v-belt at the bottom of the instrument is not loosened after using the
instrument for 6-12 months. Press the middle part of the v-belt inward about 10 mm.
Please check the abrasion condition of the v-belt after using the instrument for 2 years
and change the same type of v-belt (Z1000) if necessary.
●
Please disassembly the side plate at the right and back of the instrument to clean out
dust on the condenser and fan to ensure good heat dissipation and high cooling
efficiency.
●
Dehumidification is required if the humidity of the instrument is too high.
●
Ensure good ventilation of the room.
●
Please make sure the air circulation is sufficient when using CO
2
. The CO
2
concentration will increase in a small confined room and do harm to your health. Also,
please avoid direct inhalation of the CO
2
gas when opening the the incubator door.
The instrument should be palced in a clean and dry environment. Any heat source or
7. Structure
Temperature and
speed control panel
CO
2
control panel
Handle
Ventilation hole
Adjustable footings
Infrared lamp switch
Power switch
Fluorescent lamp
switch
Filter
CO2 switch
Power inlet