background image

PU Pipe
              outer dia. 10mm)

(

Inner dia.7mm,

Warning

9

10

5. Installation

5. Installation

Lifting the instrument by hand should be prohibited. Please use a lift or a forklift or other
lifting apparatus to raise or move the instrument.

Note

Note

5.1 Tools Requirment

a. 2 adjustable wrenches
b. 2 cross screwdriver
c. 1 spirit level
d. 1 lift or forklift or other lifting apparatus with the minimum
    loading to be 250 kg

5.2 Installation

a. Place the instrument on the ground as
    instructed in Chapter 3.1;
b. Adjust footings: as shown in the figure,
    1 of the 4 footings is adjustable.
c. Repeat step b to stack instruments.

Footing Adjustment Method:
1.After stacking the the instrument, turn the hexagon nut down using the wrench to loosen.
2.Insert the screwdrive into the adjustment hole and turn the footing. After the instrument the
leveled (the 4 footings are at the same level), turn the hexagon nut up to tighten.

Please make sure each instrument is level and not shaking when stacking multiple of them.
It is forbidden to load the instrument more than 15 kg of the maximum load. It is for bidden
to exceed the maximum speed, especially for instruments at the 2nd and 3rd levels.

Connect CO

2

 gas tank

1. Install a regulator with flow control valve onto the gas tank (see optional accessories

2

table). Please choose a regulator with the primary preset pressure to be 25 MPa (

250kg/cm

.

2

2

3500l b/in ), and secondary preset pressure to be 0.6 MPa (

2.0kg/cm

30psiG

).

2. Using the provided pipe to connect the regulator
with the instrument. The gas inlet port is
positioned at the right side of the instrument
(as shown in figure 1).

3. Adjust the secondary flow control valve to point the pointer to 0.1

– 0.2MPa (as shown in figure 2). It may cause the pipe connect the

incubator and the inlet port to fall off and the leak of the CO

2

 gas if the

secondary pressure is too high. High CO

2

 concentration will do harm 

to your health and may cause suffocating and death. Maintenance is
required if the inner pipe falls off.

4. Check if there is leak at the connections (pipe with regulator, pipe with
 inlet port of the instrument, and pipe with filter).

5. Connection components

    A. Schematic diagram

B. Actual picture

Gas Inlet

PU Pipe

 

F

ilter

Protection Cover

Figure 1

Figure 2

To the filter

Stainless steel hoop

To the CO

regulator

Note: 

1. Do not turn the pressure adjustment rod on the regulator when changing the gas tank.
It will change the secondary pressure if the pressure adjustment rod is turned and more
adjustment will be needed. It may cause the pipe connect the incubator and the inlet port
to fall off. Maintenance is required if the inner pipe falls off.
2. Inspect the conncection of the pipe regularly to ensure it is safely connected. If the pipe
is damaged or aged, change pipe.
3. Use the stainless steel hoop to tighten the connection point of the PU pipe and the filter
to prevent the pipe from falling off and leakage of gas.
4. The CO

2

 gas purity is more than 99.99%. Low gas purity may cause blocking of the

filter or damage of the electromagnetic valve.

Make sure the gas supply meets the specified requirements. Make sure all the
pipes are properly connected. Make sure the gas pressure settings are correct.
Wrong pipe connection or pressure settings may cause CO

2

 leakage. High CO

2

concentration will do harm to your health and may cause suffocation and death.
Alternative methods should be taken if there is no sufficient ventilation in the
room to guarantee safety. These methods include gas monitoring and alarming
system. Maintain correct gas pressure can prevent gas leakage.

Hexagon Nut

Summary of Contents for IncuShaker LOM-7450-CO2

Page 1: ...Stackable Incubator Shaker User s Manual Please read this user s manual carefully before using the product LOM 7450 7450 CO2 7450 L...

Page 2: ...choosing IncuShaker Series Incubator Shaker Your trust and support in MRC LTD are greatly appreciated MRC Incubator Shaker Series is model designed for shaking and heat samples according to the resear...

Page 3: ...e instruments are running with the maximum load it is recommended to set the rotation speed no more than 250 rpm 11 Please make sure the vent of the instrument is kept away from other vents or heat so...

Page 4: ...rview 2 2 Features Large view window LCD dot matrix display User friendly interface Multi segment programming Extra long timing up to 999 hours Soft start and smooth acceleration Open door cut off swi...

Page 5: ...Range Platform Size Platform Number Maximum Capacity Standard Configuration Power Power Supply Overall Size Net Weight CO2 Concentration Rang CO2 Concentration Precision CO2 Detecting sensor Chamber H...

Page 6: ...X H c Overal size when stacking three units W X D X H 124 78 67cm 124 132 67cm 124 78 200cm Please leave adequate space around the instrument to help the cooling system of the instrument to dissipate...

Page 7: ...shown in figure 2 It may cause the pipe connect the incubator and the inlet port to fall off and the leak of the CO2 gas if the secondary pressure is too high High CO2 concentration will do harm to yo...

Page 8: ...tening screws on the flask clamp regularly to prevent it from falling off Please clean the instrument regularly Please check if the v belt at the bottom of the instrument is not loosened after using t...

Page 9: ...to stop the instrument when it s in Simple Operation or Timing Program mode 9 Settings Turn on the power switch at the right side of the instrument after the instrument is properly placed Thanks for...

Page 10: ...s after finishing running the program Return to Timing Program page When Timing Program Operation shows on the display window press the button to choose whether to continue running Stoppig Y N Press t...

Page 11: ...ing Main Interface Press the button to choose a function and enter the button to enter the corresponding operation page a Basic settings Press the button to choose the parameter that needs to be modif...

Page 12: ...e current parameters and return to the temperature control parameter settings C 2 2 Temperature Calibration The value outside of the bracket is the target temperature value the value inside the bracke...

Page 13: ...he calibrated value are not saved If press the button in the calibration mode the instrument will exit the CO2 calibration and the target value is not saved and reset to factory settings The instrumen...

Page 14: ...for more than 5 minutes The alarm indicator light will flash and the buzz will sound The electromagnetic gas valve will be closed automatically Press any key to cancel the alarm status and the instru...

Page 15: ...t connected Power outlet out of power Power switch off Change to a new fuse with the same specification Position at the powr outlet Position inside the electrical cabinet Remove interference source of...

Page 16: ...For Model LOM 7450 CO2 only For Model LOM 7450 CO2 only For Model LOM 7450 CO2 only For Model LOM 7450 CO2 only For Model LOM 7450 CO2 only Optional Accessories Name Specifications Name Specifications...

Reviews: