1. Unpacking Guide

1. Unpacking Guide

Note

01. Please contact the manufacturer or your local representative immediately if the

      instrument is upside down upon unpacking.

02. Inspect package and contents upon receipt of the instrument. If the instrument is

      damaged or if there are any missing accessories according to the packing list, please

      contact the manufacturer immediately.

03. Place the instrument on a level and firm surface for use to avoid vibration and noise.

04. Please make sure the power is disconnected and the instrument is not loaded when

      moving the instrument. Please move instrument individually before stacking them.

05. When stacking multiple instruments, please make sure the floor load bearing ability is

      no less than 1.5 times of the total weight of the instruments. Please place the

      instruments beside a wall or on top of a floor with reinforced beam to ensure good

      loading condition.

06. Please use the specified fuse and make sure to disconnect power when replacing the

      fuse.

07. Please unplug the instrument during maintenance.

08. Do not clean the instrument with corrosive or flammable liquid and make sure this

      type of liquid is kept away from the instrument.

09. Please use a power supply specified on the nameplate on the right side of the

      instrument.

10. The maximum load of an individual unit is 15 kg. When three units are stacked, the

      maximum speed is 250 rpm for the top unit and 300 rpm for the middle. When the

      instruments are running with the maximum load, it is recommended to set the

      rotation speed no more than 250 rpm.

11. Please make sure the vent of the instrument is kept away from other vents or heat

      sources. Otherwise, the compressor may not be able to start normally or start

      frequently during the cooling process.

12. Please make sure the gas supply type is in accordance with the requirements and

      all the pipes are correctly and firmly connected. Please set the gas pressure as

      specified.  Any wrong pipe connection or inappropriate air pressure settings may

      cause a CO

2

 gas leak. High CO

2

 concentration may harm your health and

      potentially lead to suffocation or death.

13. Please ensure sufficient ventilation when using CO

2

. CO

2

 concentration will

      increase in a small confined room and high CO

concentration is harmful to health.

      In addition, please avoid direct inhaling of CO

2

 when opening the compartment door

      of the product.

14. Please place the 4 small footings of the instrument on

      top of the 4 big fixed footings when stacking multiple

      units.

Lifting the instrument by hand should be prohibited. Please use a lift or a forklift or other
lifting apparatus to raise or move the instrument.

Warning

Unpacking Procedures

1. Place the instrument on a level surface with caution. Remove the packing material and

    inspect if the instrument is damaged.

2. Please find the packing list, in the user manual, in the plastic documents pouch inside

    the packing box.

3. Check the packing list in the manual make sure there are no missing accessories.

    Otherwise, please contact the manufacturer immediately.

push

step 1

step 2

step 3

step 1

step 2

step 3

Summary of Contents for IncuShaker LOM-7450-CO2

Page 1: ...Stackable Incubator Shaker User s Manual Please read this user s manual carefully before using the product LOM 7450 7450 CO2 7450 L...

Page 2: ...choosing IncuShaker Series Incubator Shaker Your trust and support in MRC LTD are greatly appreciated MRC Incubator Shaker Series is model designed for shaking and heat samples according to the resear...

Page 3: ...e instruments are running with the maximum load it is recommended to set the rotation speed no more than 250 rpm 11 Please make sure the vent of the instrument is kept away from other vents or heat so...

Page 4: ...rview 2 2 Features Large view window LCD dot matrix display User friendly interface Multi segment programming Extra long timing up to 999 hours Soft start and smooth acceleration Open door cut off swi...

Page 5: ...Range Platform Size Platform Number Maximum Capacity Standard Configuration Power Power Supply Overall Size Net Weight CO2 Concentration Rang CO2 Concentration Precision CO2 Detecting sensor Chamber H...

Page 6: ...X H c Overal size when stacking three units W X D X H 124 78 67cm 124 132 67cm 124 78 200cm Please leave adequate space around the instrument to help the cooling system of the instrument to dissipate...

Page 7: ...shown in figure 2 It may cause the pipe connect the incubator and the inlet port to fall off and the leak of the CO2 gas if the secondary pressure is too high High CO2 concentration will do harm to yo...

Page 8: ...tening screws on the flask clamp regularly to prevent it from falling off Please clean the instrument regularly Please check if the v belt at the bottom of the instrument is not loosened after using t...

Page 9: ...to stop the instrument when it s in Simple Operation or Timing Program mode 9 Settings Turn on the power switch at the right side of the instrument after the instrument is properly placed Thanks for...

Page 10: ...s after finishing running the program Return to Timing Program page When Timing Program Operation shows on the display window press the button to choose whether to continue running Stoppig Y N Press t...

Page 11: ...ing Main Interface Press the button to choose a function and enter the button to enter the corresponding operation page a Basic settings Press the button to choose the parameter that needs to be modif...

Page 12: ...e current parameters and return to the temperature control parameter settings C 2 2 Temperature Calibration The value outside of the bracket is the target temperature value the value inside the bracke...

Page 13: ...he calibrated value are not saved If press the button in the calibration mode the instrument will exit the CO2 calibration and the target value is not saved and reset to factory settings The instrumen...

Page 14: ...for more than 5 minutes The alarm indicator light will flash and the buzz will sound The electromagnetic gas valve will be closed automatically Press any key to cancel the alarm status and the instru...

Page 15: ...t connected Power outlet out of power Power switch off Change to a new fuse with the same specification Position at the powr outlet Position inside the electrical cabinet Remove interference source of...

Page 16: ...For Model LOM 7450 CO2 only For Model LOM 7450 CO2 only For Model LOM 7450 CO2 only For Model LOM 7450 CO2 only For Model LOM 7450 CO2 only Optional Accessories Name Specifications Name Specifications...

Reviews: