or other lifting apparatus to raise or move the instrument.
●
The CO2 concentration may increase and do harm to your health if using the
instrument in a small confined room.
Please ensure sufficient ventilation when using the instrument with CO2 function on.
The gas concentration will increase in a small confined room and high concentration
of CO2 is dangerous for human health. Additionally, please avoid direct inhalation of
the CO2 gas when opening the chamber door.
Lifting the instrument by hand should be prohibited. Please use a lift or a forklift
7
8
3. Preparation
The instrument must be placed on a firm and level surface. The total weight of the
instrument when fully loaded must be taken into consideration. There must be suffient
space left around the instrument. It should be at least 50 cm between the instrument
and the wall or any two adjacent instruments.
3.1 Placement Conditions
Warning
3.2 Ambient Conditions
a. Environmental temperature condition:
b. Humidity condition:
10
℃~
35
℃
20
~
80% RH
3.3 Space Conditions
a. Instrument size with door closed(W X D X H):
b. Instrument size with door open(W X D X H):
c. Overal size when stacking three units(W X D X H):
124×78×67cm
124×132×67cm
124×78×200cm
Please leave adequate space around the instrument to help the cooling system of the
instrument to dissipate heat and future maintenance. It should be at least 50cm between
the instrument and the wall or any two adjacent instruments.
3.4 Power Supply Conditions
Please use the power supply voltage as listed on the nameplate at the right side of the
instrument.
4. Safety Instructions
Lifting the instrument by hand should be prohibited. Please use a lift or a forklift or other
lifting apparatus to raise or move the instrument.
●
Make sure whether the local power supply voltage is suitable for use.
●
The instrument must be properly grounded.
●
Use independent power supply.
●
Do not use extended power cord.
●
Do not drag the power cord when unplugging.
●
Do not use non-specified nor damaged power cord.
●
Only professional personnel are allowed to open the power box.
●
Tighten the fixing handle on the aluminium platform.
●
Ensure good ventilation of the room.
●
Please make sure to follow the safety instructions!
The power supply must be unplugged under any of the situations below:
When moving the instrument
When opening the power box
When changing the fuses
When the instrument is malfunctioning
When the instrument is not in use
●
●
●
●
●