background image

● 

    or other lifting apparatus to raise or move the instrument.

● 

The CO2 concentration may increase and do harm to your health if using the

    instrument in a small confined room. 

Please ensure sufficient ventilation when using the instrument with CO2 function on. 

The gas concentration will increase in a small confined room and high concentration

of CO2 is dangerous for human health. Additionally, please avoid direct inhalation of

the CO2 gas when opening the chamber door.

Lifting the instrument by hand should be prohibited. Please use a lift or a forklift

7

8

3. Preparation

The instrument must be placed on a firm and level surface. The total weight of the

instrument when fully loaded must be taken into consideration. There must be suffient

space left around the instrument. It should be at least 50 cm between the instrument

and the wall or any two adjacent instruments.

3.1 Placement Conditions

Warning

3.2 Ambient Conditions

a. Environmental temperature condition: 

 

b. Humidity condition: 

10

℃~

35

20

80% RH

3.3 Space Conditions

a. Instrument size with door closed(W X D X H): 
b. Instrument size with door open(W X D X H): 
c. Overal size when stacking three units(W X D X H): 

124×78×67cm

124×132×67cm

124×78×200cm

Please leave adequate space around the instrument to help the cooling system of the
instrument to dissipate heat and future maintenance. It should be at least 50cm between
the instrument and the wall or any two adjacent instruments.

3.4 Power Supply Conditions

Please use the power supply voltage as listed on the nameplate at the right side of the

instrument.

4. Safety Instructions

 Lifting the instrument by hand should be prohibited. Please use a lift or a forklift or other

        lifting apparatus to raise or move the instrument.

 Make sure whether the local power supply voltage is suitable for use.

 The instrument must be properly grounded.

 Use independent power supply.

 Do not use extended power cord.

 Do not drag the power cord when unplugging.

 Do not use non-specified nor damaged power cord.

 Only professional personnel are allowed to open the power box.

 Tighten the fixing handle on the aluminium platform.

 Ensure good ventilation of the room.

Please make sure to follow the safety instructions!

The power supply must be unplugged under any of the situations below:

 When moving the instrument

 When opening the power box

 When changing the fuses

 When the instrument is malfunctioning

 When the instrument is not in use





Summary of Contents for IncuShaker LOM-7450-CO2

Page 1: ...Stackable Incubator Shaker User s Manual Please read this user s manual carefully before using the product LOM 7450 7450 CO2 7450 L...

Page 2: ...choosing IncuShaker Series Incubator Shaker Your trust and support in MRC LTD are greatly appreciated MRC Incubator Shaker Series is model designed for shaking and heat samples according to the resear...

Page 3: ...e instruments are running with the maximum load it is recommended to set the rotation speed no more than 250 rpm 11 Please make sure the vent of the instrument is kept away from other vents or heat so...

Page 4: ...rview 2 2 Features Large view window LCD dot matrix display User friendly interface Multi segment programming Extra long timing up to 999 hours Soft start and smooth acceleration Open door cut off swi...

Page 5: ...Range Platform Size Platform Number Maximum Capacity Standard Configuration Power Power Supply Overall Size Net Weight CO2 Concentration Rang CO2 Concentration Precision CO2 Detecting sensor Chamber H...

Page 6: ...X H c Overal size when stacking three units W X D X H 124 78 67cm 124 132 67cm 124 78 200cm Please leave adequate space around the instrument to help the cooling system of the instrument to dissipate...

Page 7: ...shown in figure 2 It may cause the pipe connect the incubator and the inlet port to fall off and the leak of the CO2 gas if the secondary pressure is too high High CO2 concentration will do harm to yo...

Page 8: ...tening screws on the flask clamp regularly to prevent it from falling off Please clean the instrument regularly Please check if the v belt at the bottom of the instrument is not loosened after using t...

Page 9: ...to stop the instrument when it s in Simple Operation or Timing Program mode 9 Settings Turn on the power switch at the right side of the instrument after the instrument is properly placed Thanks for...

Page 10: ...s after finishing running the program Return to Timing Program page When Timing Program Operation shows on the display window press the button to choose whether to continue running Stoppig Y N Press t...

Page 11: ...ing Main Interface Press the button to choose a function and enter the button to enter the corresponding operation page a Basic settings Press the button to choose the parameter that needs to be modif...

Page 12: ...e current parameters and return to the temperature control parameter settings C 2 2 Temperature Calibration The value outside of the bracket is the target temperature value the value inside the bracke...

Page 13: ...he calibrated value are not saved If press the button in the calibration mode the instrument will exit the CO2 calibration and the target value is not saved and reset to factory settings The instrumen...

Page 14: ...for more than 5 minutes The alarm indicator light will flash and the buzz will sound The electromagnetic gas valve will be closed automatically Press any key to cancel the alarm status and the instru...

Page 15: ...t connected Power outlet out of power Power switch off Change to a new fuse with the same specification Position at the powr outlet Position inside the electrical cabinet Remove interference source of...

Page 16: ...For Model LOM 7450 CO2 only For Model LOM 7450 CO2 only For Model LOM 7450 CO2 only For Model LOM 7450 CO2 only For Model LOM 7450 CO2 only Optional Accessories Name Specifications Name Specifications...

Reviews: