MRC ASB-110-X Operation Manual Download Page 66

- 66 - 

 

 

A label example named  „Etykieta01” 
 

 

 
3 labels after each measurement are to be printed out 
 
Printout designing procedure
 

 

Inscribe printout data in corresponding texts – P6 parameters’ 
group Printout; parameters: String 01 

÷

 String 80. 

 
Use variables from chart 1 (printout control variables) and variables 
including different data from the balance. 
 
Table 1 

 

\02L\c 

Start of the label 

rlabelname\c 

Label name 

X\c 

Start of variables’ edition 

\02U01NN\c 

Variable 1 NN-variable symbol 

\02U02NN\c 

Variable 2 NN-variable symbol 

\02U03NN\c 

Variable 3 NN-variable symbol 

\02UnnNN\c 

Variable nn NN-variable symbol 

\02fnnn\c 

Paper feeding nnn [mm] – depending on the label size 

E\c 

End of variables’ edition 

\02Ennnn\c 

Print nnn labels 

\02G\c 

End of the label 

 

Summary of Contents for ASB-110-X

Page 1: ......

Page 2: ...2 JULY 2010...

Page 3: ...16 5 2 User menu 16 6 WEIGHING 17 6 1 Log in function 18 7 BALANCE CALIBRATION 21 7 1 Automatic balance calibration 21 7 2 Calibration test 23 7 3 Manual calibration 24 7 3 1 Internal calibration 24 7...

Page 4: ...ity of solids and liquids 47 16 6 1 Density of liquids 47 16 6 2 Density of solids 48 16 7 Formulation 48 16 8 Statistics 53 17 KINDS OF PRINTOUTS 55 17 1 Standard printout 55 17 2 Non standard printo...

Page 5: ...amber place lower shield of weighing chamber 1 inside the chamber place anti breeze shield 2 put centring ring on the shield 3 put the balance pan inside the ring 4 close side doors of weighing chambe...

Page 6: ...balance should be replaced in leveled position 2 2 Set level up The balance should be leveled The correct leveling is shown on the level indication installed at rear of the balance 2 3 Warming up Befo...

Page 7: ...counts pieces weights dosages determines deviations of the standard mass weights animals detemines liquids and solids density formulation creates statistics Measure units and particular functions can...

Page 8: ...ESCRIPTION OF THE BALANCE display Draw 3 Display balance 1 load mass and quantity of pieces 2 measure unit 3 the result is stable 4 line of max range of the balance 5 work mode 6 date 7 time 8 precise...

Page 9: ......

Page 10: ...d contents of each group is presented below P1 Calibration 01 Int calibr function 02 Ext calibr function 03 User calibr function 04 Calibr test function 05 Weight corr 0 0 06 Auto calibr 0 3 both 07 A...

Page 11: ...05 Handsake 0 none 06 Auto print 0 none 07 Interval 1 0 1 s 08 Min mass 4 10 d 09 Print on stab 1 enabled 10 Printer type 0 Epson Standard 11 Cut paper 0 no P6 Printout 01 Printout No 0 standard 02 Pr...

Page 12: ...ting function 02 ID autoprint 0 off 03 Beep 1 enabled 04 Language 1 english 05 Backlight 1 on 06 Contrast function 07 Screensaver 0 enabled 08 Temperature function 09 Balance Id 114493 10 Software rev...

Page 13: ...ng the RIGHT ARROW key display II Draw 5 Menu 1 main menu number 2 indication of the function selection 3 function name 4 currently used function 5 submenu number 6 submenu name 7 attribute of the men...

Page 14: ...is presented on the display move one level up to main menu Resignation parameter changing 5 1 2 Return to weighing function Introduced changes are recorded after return to weighing mode and confirm c...

Page 15: ...d has its equvivalent on the PS keyboard for functions Description keyboard Switch on off the balance display Move to the balance menu Selects work mode Selects measure unit PRINT TARE for direction k...

Page 16: ...lance manu and control the balance work The list of the commands is at the end of the manual 5 2 User menu The menu is presented in p 5 Press the F key in weighing level Main menu is presented on the...

Page 17: ...Stable temperature Stable ground Proper parameters for external conditions Before measurements or for essential changes of the external conditions if the temperature changes more than 1 C h calibrate...

Page 18: ...abilization 6 1 Log in function The operator has own access code to internal menu The password system is defined by administrator Password can contain max 6 digits The balance program enables to decla...

Page 19: ...ected by password Start up If it is settled on YES during start the balance up user must write access password administrator or user Functions If it is settled on NO not protected by password user can...

Page 20: ...n Please remember the password If you set YES for Start up function the password must be writen after switch the balance on If the password is not correct using the balance is not possible To write th...

Page 21: ...rd Calibration made with external weight with declared mass which cannot be modificated with any mass which should be given before the calibration process In verified balances only automatic internal...

Page 22: ...m changes on the drawing field No 9 and No 10 01 Internal calibration Start internal calibration process the process is automaticaly without operator interference if there is load on the pan the displ...

Page 23: ...c calibration time Determination of time automatic calibration starts up Return to weighing The changes are recorded when the balance returns to weighing mode with the recording the changes Press the...

Page 24: ...tion in the memory and returns to weighing mode Pressing the ESC key stops the calibration process If during the calibration load is on the pan display show order about error The calibration process i...

Page 25: ...ance displays order to note calibration mass The first digit flashes and it can be changed Draw 14 User calibration declaring value of weight Record new external mass by functional keys in accordance...

Page 26: ...the report is genrated and sent automaticaly Draw 15 Submenu calibration A content of report depends on setting in submenu GLP All options with YES attribute are printed Draw 16 Submenu GLP setting A...

Page 27: ...by the balance keyboard or the PS 2 keyboard For the rest select 1 no do not print during report 0 yes print during report Main view of the GLP submenu is presented on drawing 19 page 24 If you use th...

Page 28: ...e form 24 h After selection press the ENTER to confirm 12 h form is distinguished by the letters PM or AM on the printouts 03 Time Enter setting the parameter 03 Time by the RIGHT ARROW key in accorda...

Page 29: ...key The balance returns to submenu P3 Date Time and displayed time changes After setting time return to weighing mode in accordance to p 5 1 1 of the manual 04 Date Set the parameter 04 Date with the...

Page 30: ...ans of parameters in group P4 Readout Draw 22 Submenu Readout internal setting 10 1 Setting filter Dependly on conditions set the filter If the conditions are conductive set the filter as very fast va...

Page 31: ...ontrols automaticaly and corrects zero indication of the balance If the function is active following results in declared periods of time are compared eg each 1s If these results differs at less value...

Page 32: ...2 9600 3 19200 07 Interval Interval it is defined how often balance sends indications through RS 232 port It is counted on base on form for the parameter x 0 1 s time yestu interval Value from 1 to 99...

Page 33: ...s user declares mass units which are accessible for operator directly under the key Units All units which value of the parameters is set up at 1 yes are accessible from the level of switching between...

Page 34: ...the parameters can be done according to p 5 1 1 of the manual 15 OTHER PARAMETERS User can set parameters have influence on work with balance in group of the paraemeters P9 Others eg beep signals etc...

Page 35: ...of the measurement does not change 08 Temperature it is information about temperature which is registered by temperature sensor in the balance Return to the menu press the ESC key 09 The balance numbe...

Page 36: ...fast procedure of introducing new setting After printing actual parameters to file in the computer user changes the parameters very fastly and without any problems User sends new corrected setting to...

Page 37: ...he ENTER or select 07 Start and press the RIGHT ARROW Functions to count details are activated Draw 28 Counting pieces display view APW singular piece mass g WGH all pieces mass on the pan pcs mark fo...

Page 38: ...the key RIGHT ARROW In the counting pieces function press the key Units Dialog window appears on the display Select the batch quantity fiels 01 04 or write it in 05 Standard Draw 29 Countimg pieces wi...

Page 39: ...r adding must be higher than it was previously 2 quantity of pieces after adding must be less than twice quantity which was on the display before adding 3 actual quantity must be in tolerance 0 3 of t...

Page 40: ...e Draw 32 Select piece from date base Select piece form date base Start counting pieces 16 2 Checkweighing The sample is weighed precisely when the limits of weighing are settled The process is shown...

Page 41: ...5 value of low LO and HI high limit 6 graphs which presents weighign range Rememeber to set the parameter 02 High limit firstly The balance program checks if the values are correct and if they are in...

Page 42: ...he procedure set the standard mass which is upper stage of the dosage Activation of the function Draw 35 Dosage activation of the function Display Draw 36 Dosage display 1 mass which should be added 2...

Page 43: ...d in percentages Following functions dosage weighing statistics can cooperate with deviation function Activation of the function Draw 37 Percents activation of the fucntion Display Draw 38 Percents di...

Page 44: ...Statistics select field Cancel and cancel previous statistics and change the attribute NO into attribute YES Confirm this option and press the key Enter Draw 39 Percents cooperation with ther function...

Page 45: ...the parameter 05 Statistics Pressing the F key and enter the parameter 05 Statistics Select the parameter 02 Results Enter function of showing statistics results After pressing the ENTER statistics r...

Page 46: ...r filter the shorter time of measurement STAGE Value in actual scale intervals is value the result must be below The result of weighing must be smaller than value of actual scale intervals in order to...

Page 47: ...r 12 Attachments 16 6 1 Density of liquids Basic component during measure solids of liquids is glass float It has precise determined capacity which is stamped on the float hook Write password to balan...

Page 48: ...tting the key PRINT Precise description of measurements performance and setting is the manual of Specific Gravity Measurement Kit 16 7 Formulation This function is used to make mixtures under recipes...

Page 49: ...ter confirmation mas sof each component parameter 03 Quantity of components user determines quantity of components the mixture should include max 20 signs parameter 04 Recipe after set this parameter...

Page 50: ...y for recipes Draw 44 Recipes functions 1 mass which is actualy on the pan 2 stable measurement sing 3 function name 4 settled mass of the weighed component in the parameter 04 Recipe 5 Sum of weighed...

Page 51: ...balance there cannot be more than 20 components in the mixture Write quantity of components in the parameter 03 Quantity of components Quantity of components cannot be higher than 20 pcs The program r...

Page 52: ...nts and their mass date in the balance memory If documentation of preparong mixture is printed set the parameter 02 Printout at 1 YES If mass of each component is confirmed each mass with their names...

Page 53: ...int results press the key PRINT 16 8 Statistics Activation Draw 47 Statistics fucntion activation Results of previous statistics should removed after function activation It is realized through option...

Page 54: ...8 g max mass D 15 092 g difference between Max Min SDV 6 581 g standard deviation RDV 20 4 variation factor Draw 48 Statistics display for series of measurement 1 mass on the pan 2 measurement number...

Page 55: ...printout It includes result of weighing and all variables which have attibute YES in GLP submenu In User and Project fields names should be written Draw 49 Declaration of variables to printout submenu...

Page 56: ...exts from 1 to 40 are printed And then write text in the lines 1 40 It is recomended to use PC keyboard what is simplier and faster way Non standard printouts can overlap each other Printout 1 Start 1...

Page 57: ...texts 8 signs each from the parameter Text 1 to Text 80 User can design 4 non standard printouts 17 2 1 Texts Variables in all modes and with the same values Printout of singular sign N Actual net ma...

Page 58: ...p Procedure c Temperature a Density of liquid v Float capacity MEASUREMENT OF DENSITY Statistic variables in all modes apart from basic weighing n The measurement number x Average value S Sum m Min v...

Page 59: ...xts User can use special signs to include variables dependly on own necessities Example 1 Max mass cannot be higher than 11 250 g If user write this sentence uses 640 signs grouped in adjacent lines o...

Page 60: ...Text Parameter number 1 2 3 4 5 6 7 8 25 Text 16 Z a k a d M 26 Text 17 e c h a n i k i 27 Text 18 P r e c y z y 28 Text 19 j n e j R A 29 Text 20 D W A G c D 30 Text 21 a t a d c 31 Text 22 G o d z i...

Page 61: ...to change and move lef after this key is pressed flashing sign is cancelled Confirm the text On PS 2 keyboard Press F2 to enter main menu Press F3 to set parameters indications next to group P6 Print...

Page 62: ...62 17 2 2 Composing texts by Edition function The function activation Draw 53 Not standard printouts printout edition Draw 53 1 Not standard printouts printout edition select...

Page 63: ...keys ARROWS TOP UP To print field press the RIGHT ARROW key After the edition press ENTER PRINT Display shows question if printout should be done press ENTER PRINT again 17 2 3 Select non standard pr...

Page 64: ...o the computer or printer press the PRINT key 9600 bit s is the default setting for the speed of transmission If the peripheral unit has a different speed of transmission you may change the speed of t...

Page 65: ...abel Notice In order to print properly interial inside balance variable you should predict a proper number of characters for each variable charts 2 3 and 4 After giving it an original name save the la...

Page 66: ...control variables and variables including different data from the balance Table 1 02L c Start of the label rlabelname c Label name X c Start of variables edition 02U01NN c Variable 1 NN variable symb...

Page 67: ...weighing mode save the parameters Now you can connect the balance with the printer a cable diagram in the users manual Check the setting of transmition parameters both in the balance and in the print...

Page 68: ...number R 8 Program number P 8 Project number U 8 Operator number F X Name of present operation mode C 25 Date and time of the last calibration K X Type of the last calibration I 16 or 18 Deviation in...

Page 69: ...ls weighing i 14 Liquid p 14 Procedure c 14 Temperature a 16 Liquid mass density v 16 Plunger volume Mass density measurement Table 4 Statistical variables that are present in every operation mode exc...

Page 70: ...ce There is a hook visible which can be used to suspend a sample Note 1 The suspension eye is directly connected to the weighing mechanism Take great care not to move or rotate the eye as this will da...

Page 71: ...MANDS FROM THE BALANCE Command PC CR LF all recorded information in commands in the balance programme are sent from the balance Function SEND THE RESULT IN BASIC UNIT Command S CR LF result is sent fr...

Page 72: ...WRITE OR CHANGE DATE IN THE BALANCE Command PD CR LF the balance sends settled date or the date is changed Function WRITE NEW OR CHANGE TIME IN THE BALANCE Command PD CR LF the balance sends settled...

Page 73: ...KEYBOARD Command KL CR LF Function UNBLOCK THE KEYBOARD Command KU CR LF Function SWITCH ECHO OFF FOR THE KEYBOARD Command E0 CR LF keys codes are switched off Function SWITCH ECHO ON FOR THE KEYBOARD...

Page 74: ...1 Exceed max measure range of the balance Exceed range 2 2 Exceed max measure range of the balance A D Null 2 3 No divisions from converter A D Full 2 4 Exceed max value converter intervals Tara Zero...

Page 75: ...ove the range 3 1 The parameter value above the range Faulty value 3 2 Inadmissible value of the paraemter Blocked DRH 3 3 The parameter cannot be changed Function DRH active in the factory menu Writt...

Reviews: