- 23 -
05 Weight code
correct value of internal calibration mass, function inadmissible in
verified balances
06 Automatic calibration
determine factor which decides about start automatic internal calibration
0
non – non of the factors causes start of the calibration
1
time – calibration in relation to time determined in p. 07
2
temperature – calibration in relation to changes of surroundign
temperature
3
both – calibration in relation to changes of time and temperature
07 Automatic calibration time
Determination of time automatic calibration starts up
Return to weighing
The changes are recorded when the balance returns to
weighing mode with the recording the changes. Press the
ESC many times. Following question appears on the display.
Select one of the options : ENTER – record / ESC – cancell
(see. Draw 9. Return to weighing p. 5.1.2. Return to weighing)
7.2. Calibration test
Internal calibration mass is compared to its value in the balance memory.
This preocess is automatic. Its result is shown on the display.
Draw 13. The calibration test
Cal. – value of internal calibration mass
Act. – result of weighing internal calibration mass
Dif – difference between two values