- 22 -
calibration process is stopped.
Set up automatic calibartion
Draw 12. Set up automatic balance calibration
1 – main menu number
2 – function selection factor
3 – function name
4 – name of actual activity
5 – selects factor to autocalibration (time / temp.)
6 – declaring autocalibration time
7 – value of factors for autocalibration
8 – value of time for autocalibration
if the
values for factor and autocalibration time also descriptions for tchem
changes (on the drawing field No 9. and No 10.)
01 Internal calibration
Start internal calibration process, the process is automaticaly without
operator interference, if there is load on the pan the display shows
order to remove the load
02 External calibration
calibration performed by external mass, its value is recorded in factory
menu, function inadmissible in verified balances
03 User calibration
calibration performed with any mass which must be introduced before
the calibration, function inadmissible in verified balances
04 Calibration test
comparison internal calibration mass with its value recorded in the
balance memory