8.6 Capacity measurement
Before performing capacity measurements, discharge capaci
‑
tors completely!
Never apply voltage to the capacity measurement jacks!
This might damage or destroy the device! A damaged device
might represent an electrical hazard!
- Select the desired function (
) by means of the rotary switch
8
of the
01
- Determine the polarity of the capacitor and completely discharge the
capacitor.
- Connect the black safety measuring line to the COM jack
J
of the
01
- Connect the red safety measuring line to the jack for V,
,
, Hz
9
of the
01
- Bring the safety measuring lines into contact with the discharged capacitor
according to ist polarity and read the measured value on the digital display
of the
01
See figure 9:
Capacity measurement
8.7 Frequency measurement
- Select the desired function (Hz) by means of the rotary switch
8
of the
01
- Connect the black safety measuring line to the COM jack
J
of the
01
- Connect the red safety measuring line to the jack for V,
,
, Hz
9
of
the
01
. Please observe the minimum sensitivity for
frequency measurements of the
01
- Bring the safety measuring lines into contact with the measuring
points and read the measured value on the digital display
of the
01
See figure 10:
Frequency measurement
8.8 Temperature measurement
- Select the desired function (°C or °F) by means of the rotary switch
8
of the
01
- Connect the temperature measuring adapter and the temperature measuring
line to the COM jack (-)
J
and to the jack for V,
,
, Hz (+)
9
observing
correct polarity.
- Arrange the end of the temperature measuring line in the vicinity of the heat
source to be monitored. read the measured value on the digital display
of the
01
See figure 11:
Temperature measurement
8.9 Voltage indicator
the voltage indicator function is possible from each position of the rotary switch.
No measuring lines are required as voltage indicator (non-contact detection of
an alternating field). The detector is located on the upper part of the device
behind the leD. By pressing the „VoltSensor“ key
5
, the display indication
disappears (if the display is switched on). if a phase voltage is localized, this is
indicated by an acoustic signal and a red leD signal
M
. An indication is made
in earthed AC current networks only! the phase can be determined by means of
a single-pole measuring line.
Practical hint:
interruptions (cable breaks) in cables lying around openly such as e.g. cable
reels, fairy lights etc. can be traced from the feeding point (phase) to the point
of interruption.
Functional range:
≥
230 V
See figure 12:
Voltage indicator with buzzer
8.9.1 Phase test
- Connect the red safety measuring line to the jack for V,
,
, Hz
9
of the
01
.
- Bring the safety measuring line into contact with the measuring point of the
system part and press the „VoltSensor“ key
5
.
- if the red leD lights and if there is an acoustic signal, the phase of an
earthed AC voltage is applied to this measuring point (system part).
9. Maintenance
Before opening the
01
, make sure that
the device is free of voltage! Electrical danger!