8.2.1  Voltage measurement

-   Select the desired function (V AC) or (V DC) by means of the rotary switch 

8

 of the 

01

-   Connect  the  black  safety  measuring  line  to  the  COM  jack 

J

  of  the 

01

-    Connect the red safety measuring line to the jack for 

V, Ω, 

, Hz 

9

 of the 

01

-   Bring  the  safety  measuring  lines  into  contact  with  the  measuring 

points  and  read  the  measured  value  on  the  digital  display 

  of  the  

01

See figure 2:  

 

DC voltage measurement

See figure 3:  

 

AC voltage measurement

8.2.2  Current measurement 

01

-   Select the desired range and function (A AC) or (A DC) by means of the 

rotary switch 

8

 of the 

01

-   Connect the black safety measuring line to the COM jack 

J

 of the 

01

 

-   Connect the red safety measuring line to the jack for the 10 A range 

K

 (DC 

or AC currents up to 10 A) of the 

01

-   Bring  the  safety  measuring  lines  into  contact  with  the  measuring 

points  and  read  the  measured  value  on  the  digital  display 

  of  the  

01

See figure 4: 

 

DC current measurement 

See figure 5:  

 

AC current measurement 

8.3  Resistance measurement

-   Select  the  desired  function  (Ω)  by  means  of  the  rotary  switch 

8

  of  the 

01

-   Connect  the  black  safety  measuring  line  to  the  COM  jack 

J

  of  the 

01

-   Connect the red safety measuring line to the jack for V, 

, Hz 

9

 of the 

01

-   Bring  the  safety  measuring  lines  into  contact  with  the  measuring 

points  and  read  the  measured  value  on  the  digital  display 

  of  the  

01

See figure 6:  

 

Resistance measurement

8.4  Diode test

-   Select the desired function (

) by means of the rotary switch 

8

 of the 

01

-   Connect  the  black  safety  measuring  line  to  the  COM  jack 

J

  of  the 

01

.

-   Connect the red safety measuring line to the jack for V, 

, Hz 

9

 of the 

01

-   Bring  the  safety  measuring  lines  into  contact  with  the  diode  connections 

and  read  the  measured  value  on  the  digital  display 

  of  the  

01

-   For a standard Si diode applied in conduction direction, a conduction voltage 

between 0.400 V and 0.900 V is displayed. „000“ indicates a short-circuit 

inside the diode, „Ol“ indicates an interruption inside the diode.

-   For a diode applied in reverse direction, „Ol“ is indicated. if the diode is 

defective, „000“ or other values are indicated.

See figure 7: 

 

 Diode test

8.5  Continuity test with buzzer

-   Select the desired function (

) by means of the rotary switch 

8

 of the 

01

-   Connect  the  black  safety  measuring  line  to  the  COM  jack 

J

  of  the 

01

-   Connect the red safety measuring line to the jack for V, 

, Hz 

9

 of the 

01

-   Bring the safety measuring lines into contact with the measuring points. if the 

line resistance between the COM jack 

J

 and the jack for V, 

, Hz 

9

 falls 

below 25 

, the integrated buzzer of the 

MN 1-3 

sounds.

See figure 8: 

 

Continuity test with buzzer

0XOWLPHWUH01PSYQRWLFHSGI

www.motor-pump-ventilation.com

Summary of Contents for MN1-3

Page 1: ...RANGE MAX MIN HOLD VoltSensor MM 1 3 V V OFF Hz A A C F Hz A MAX 10A FUSED HOLD AUTO MAX MIN nmF www motor pump ventilation com 8VHU V PDQXDO 0XOWLPHWUH 01 PSYQRWLFH SGI...

Page 2: ...je figur 1c Apparatforside 1c 1 bra El ln zet ill 1c Lato anteriore apparecchio Fig 1c Voorzijde van het apparaat Rys 1c Panel przedni przyrz du Imaginea 1c Partea frontal a aparatului 1c Resim 1c Cih...

Page 3: ...i n de corriente cont nua obr 4 M en stejnosm rn ho proudu figur 4 J vnstr msm ling 4 4 bra Egyen ram m r s ill 4 Misura corrente continua Fig 4 Meten van gelijkstroom Rys 4 Pomiar pr du sta ego Imagi...

Page 4: ...6 Resistance measurement Fig 6 Mesure de r sistance Fig 6 Medici n de resistencia obr 6 M en odporu figur 6 Modstandsm ling 6 6 bra Ellen ll s m r s ill 6 Misura di resistenza Fig 6 Weerstandsmeting R...

Page 5: ...dik tor nap t s bzu kem figur 12 Sp ndingsindikator med brummer 12 12 bra Fesz lts g indik tor z mm g vel ill 12 Indicatore di tensione con cicalino fig 12 spanningsindicator met zoemer Rys 12 Wska ni...

Page 6: ...ginea 13 Schimbarea bateriilor 13 Resim 13 Batarya De i imi Bild 14 Sicherungswechsel Fig 14 Fuse replacement Fig 14 Remplacement des fusibles Fig 14 Cambio de fusible obr 14 V m na pojistek figur 14...

Page 7: ...en Rys 15 Zwijanie przewod w pomiarowych Imaginea 15 nf urarea cablului de m surat de siguran 15 Resim 15 Emniyet l m tesisat n n sar lmas 0XOWLPHWUH 01 PSYQRWLFH SGI Bild 16 Aufstellung des MN 1 3 Fi...

Page 8: ...ectrical circuits with a nominal voltage higher than 1000 V DC and 750 V AC see section 6 Ambient conditions for details The following symbols are used in this operating manual and on the 01 Warning o...

Page 9: ...ert the measuring lines into the respectively designated measuring sockets of the multimeter 3 Scope of delivery The scope of delivery of the 01 comprises 3 1 One 01 3 2 One safety measuring line red...

Page 10: ...g the key for approx 1 second AUTO is shown on the display 5 1 6 Measured value storage HOLD Press the HOLD key 7 to store the measuring result At the same time the display shows the HOLD symbol Press...

Page 11: ...V 0 5 of the measured value 2 digits 1000 VDC 200 V 100 mV 0 5 of the measured value 2 digits 1000 VDC 1000 V 1 V 0 5 of the measured value 2 digits 1000 VDC 7 2 AC voltage ranges The input resistance...

Page 12: ...igits 1 3 V 200 k 100 0 7 of the measured value 3 digits 1 3 V 2 M 1 k 1 0 of the measured value 3 digits 1 3 V 20 M 10 k 1 5 of the measured value 3 digits 1 3 V 6 For automatic range selection AUTO...

Page 13: ...temperatures only Do not permanently expose the device to sunlight Check stated nominal voltage and nominal current on the safety measuring lines Nominal voltage and current of the enclosed safety mea...

Page 14: ...ety measuring lines into contact with the measuring points and read the measured value on the digital display 1 of the 01 See figure 6 Resistance measurement 8 4 Diode test Select the desired function...

Page 15: ...M jack J and to the jack for V Hz 9 observing correct polarity Arrange the end of the temperature measuring line in the vicinity of the heat source to be monitored Read the measured value on the digit...

Page 16: ...integrated 1 5 V micro AAA batteries Battery replacement see figure 13 is required if the battery symbol 3 appears on the display 1 Proceed as follows to replace the battery Disconnect the safety meas...

Page 17: ...Robert Bosch Str 20 D 46397 Bocholt 9 6 Spare parts Fuse F 10 A 600 V 50 kA Bussmann KTK or DCM P no 748263 10 Using the suspension fixture You can store the safety measuring lines by winding them aro...

Reviews: