- Overvoltage category / installation category: IEC 60664-1/ IEC 61010-1 →
600 V category iii; 1000 V category ii,
- Contamination class: 2,
- Protection category: iP 30 DiN VDe 0470-1 ieC/eN 60529
iP 30 means: Protection against access to dangerous parts and protection
against solid impurities of a diameter > 2.5 mm, (3 – first index).
No protection against water, (0 – second index).
- Operating temperature and relative air humidity:
For operating temperatures from 0 °C to 30 °C: relative air humidity less
than 80 %,
For operating temperatures from 31 °C to 40 °C: relative air humidity less
than 75 %,
For operating temperatures from 41 °C to 50 °C: relative air humidity less
than 45 %,
- Storage temperature: the
01
can be stored at
temperatures between - 15 °C and + 60 °C (air humidity of 0 to 80 %).
During storage, the battery should be removed.
7.
Electrical specifications
Note: The measuring accuracy is specified as the sum of:
- a relative part of the measured value and
- a number of digits (i.e. counting steps of the last digit).
this measuring accuracy applies for temperatures from 18 °C to 28 °C and a
relative air humidity less than 80 %.
7.1 DC voltage ranges
The input resistance is 10 MΩ.
Measuring
range
*3
Resolution
Measuring accuracy
Overload
protection
200 mV
100 µV
± (0.5 % of the measured value + 2 digits)
1000 V
DC
2 V
1 mV
± (0.5 % of the measured value + 2 digits)
1000 V
DC
20 V
10 mV
± (0.5 % of the measured value + 2 digits)
1000 V
DC
200 V
100 mV
± (0.5 % of the measured value + 2 digits)
1000 V
DC
1000 V
1 V
± (0.5 % of the measured value + 2 digits)
1000 V
DC
7.2 AC voltage ranges
The input resistance is 10 MΩ in parallel 100 pF.
Measuring
range
*3
Resolution
Measuring accuracy *
1
within the
frequency range 50 Hz - 300 Hz
Overload
protection
200 mV
100 µV
± (2.0 % of the measured value + 5 digits) *
2
750 V
eff
2 V
1 mV
± (1.5 % of the measured value + 5 digits) *
2
750 V
eff
frequency range 50 Hz - 500 Hz
20 V
10 mV
± (1.5 % of the measured value + 5 digits) *
2
750 V
eff
200 V
100 mV
± (1.5 % of the measured value + 5 digits) *
2
750 V
eff
750 V
1 V
± (1.5 % of the measured value + 5 digits) *
2
750 V
eff
the measured value of the
01
is obtained by mean
value rectification and is displayed as r.m.s. value.
*1
The measuring accuracy is specified for a sinusoidal curve. In case of non
sinusoidal curves, the accuracy of the displayed value decreases. thus, an
additional error results for the following crest factors:
Crest factor from 1.4 to 3.0 additional error ± 1.5 %
Crest factor from 3.0 to 4.0 additional error ± 3 %
*2
Applies to sinusoidal curves of 50 Hz/ 60 Hz
*3
For automatic range selection (AutO), the change-over point might be
already at a value of 1400!
7.3 DC current ranges
Overload protection:
- 10 A (600 V) fuse, quick-acting, 50 kA, at the 10 A input
Measuring
range
Resolution
Measuring accuracy
Voltage drop
2 A
1 mA
± (1.0 % of the measured value + 3 digits)
max. 2 V
10 A
*5
10 mA
± (1.0 % of the measured value + 3 digits)
max. 2 V
0XOWLPHWUH01PSYQRWLFHSGI