mSquad User’s Guide 

14 

 

 
 
 

5.2

 

Man-Down Function 

 

5.2.1 Man-Down On/Off 

The Man-Down On/Off function has 4 settings: 

• 

Off 

• 

On

    Alarm is triggered with insufficient movement in any direction. The pre-warn or 

alarm can be cleared by quickly tilting, or by pressing the Left [+/OK] key. 

• 

Vertical Off

    The alarm stays off as long as the instrument is held in a vertical position, 

such as clipped to a belt (or straight upside-down), and triggers when in any other 
position including held sideways or laying face up or down. A pre-warn can be cleared by 
quickly returning the POLI to a vertical position, or pressing the Left [+/OK] key. 

• 

Horizontal Off

    The alarm stays off as long as the instrument is held in a horizontal 

position, such as laying face up on a table, and triggers when in any other position. A pre-
warn can be cleared by quickly returning the POLI to a horizontal position, or pressing 
the Left [+/OK] key. 

The Man-Down icon on the main screen verifies that the   
Man-Down function is enabled.

 

 

5.2.2

 

Man-Down Warn Time 

This menu allows adjustment of the time allowed for the user to clear a pre-warn signal before 
the unit goes into full Man-Down Alarm. The pre-warn time can be adjusted between 10 and 60 
seconds and the default value is 30 seconds. 

5.2.3

 

Man-Down Threshold (Sensitivity) 

This menu allows adjustment of the sensitivity to motion (On Mode) or position changes 
(Vertical or Horizontal Off Modes) for detecting a Man-Down condition.    Low sensitivity 
means fewer alarms and High sensitivity means a greater likelihood of triggering an alarm. 

5.2.4

 

Man-Down Motionless Time 

This menu allows adjustment of the time allowed for stopped motion (On Mode) or position 
changes (Vertical or Horizontal Off Modes), before a Man-Down condition is detected. The 
motionless or position-change time can be adjusted between 10 and 60 seconds and the default 
value is 30 seconds. 

 

 

 

 

 

 

 

 

 

Summary of Contents for mSQUAD MP400H

Page 1: ...mSQUAD Mobile Wireless Multi Gas Team Warning System MP400S MP400H User s Guide Rev 1 0 January 2020...

Page 2: ...t up 10 4 2 Alarm Notifications 10 4 3 Head Host Calibration 12 5 Configuration Mode 13 5 1 Navigating Configuration Mode to Edit Parameters 13 5 2 Man Down Function 14 5 3 Registering Soldier MP400S...

Page 3: ...RRATIC READING MAY INDICATE A GAS CONCENTRATION BEYOND UPPER SCALE LIMIT WHICH MAY BE HAZARDOUS ATTENTION DES LECTURES SUP RIEURES A L CHELLE PEUVENT INDIQUER DES CONCENTRATIONS EXPLOSIVES TOUTE LECTU...

Page 4: ...ions for Safe Use The POLI multi gas detector must be calibrated if it does not pass a bump test when a new sensor has been installed or at least once every 180 days depending on use and sensor exposu...

Page 5: ...by diffusion because the communications antenna is attached to the pump inlet port while MP400S Soldier units have a built in sampling pump for active gas sampling and thus a somewhat faster response...

Page 6: ...work for communication with mPower Suite configuration and data transfer software The mPower USB cable P N M 011 3003 000 is required for a PC to recognize the instrument and communicate with mPower...

Page 7: ...ration alarms 3 1 Turning On Off To turn on press and hold the Key for 3 seconds until the buzzer beeps and the red LED turns on To turn off in normal reading mode press and hold the key for a 5 secon...

Page 8: ...ength at 0 5 level Pump status pump versions only Datalogging enabled cannot turn off Battery voltage status Man down alarm enabled All sensors have been bump tested and calibrated and none are overdu...

Page 9: ...act as soft keys mapped to two text or symbol boxes at the LCD bottom that change numbers and make selections under various menus In addition to the functions described above the Left OK key can be u...

Page 10: ...est of connectivity can be done by sending a Panic Alarm Hold Left key for 3 seconds 4 2 Alarm Notifications The POLI provides alarms including audible buzzer visible bright LEDs vibration and text no...

Page 11: ...e alarm is silenced If a Soldier is turned off other Soldiers alarm immediately for 15 seconds The Host Head is alarmed with a symbol displayed next to the user name after the communication lost timer...

Page 12: ...ut the user is OK it can be stopped using the Left key Clear WARNING The Man Down feature cannot detect a worker in distress in all situations even if they have collapsed The Man Down function should...

Page 13: ...menus and OK Enter to enter a menu to edit the parameters in its submenu 5 1 1 Configuration Mode Menus and Sub menus Calibration Measurement Alarm Datalog Monitor Setup Wireless Exit Fresh Air Calib...

Page 14: ...ng the POLI to a horizontal position or pressing the Left OK key The Man Down icon on the main screen verifies that the Man Down function is enabled 5 2 2 Man Down Warn Time This menu allows adjustmen...

Page 15: ...e will appear on the Host Head Registered Device List with a next to the name The Host assigns default user names in the order they are registered e g USR00 USR01 USR02 To enter a personalized user na...

Page 16: ...vice Ping Assign Worker 1 05496868 JACK Ping Worker Assign Worker 1 05496868 JACK JACK Worker Assign Assign Worker 1 05496868 JACK BILL Worker Assign Worker Pings Monitor to Verify Unit Assign Worker...

Page 17: ...network 5 6 Channel Select the operating Channel from CH1 to CH16 Ensure that the same selection is made on the Head Host as on all Soldiers in the mSquad 5 7 mPlatoon Wireless Networks Future system...

Page 18: ...Edit the configuration parameters as desired including those under the tabs for each sensor in the upper right Click Write to upload the configuration to the instrument Read allows downloading the cur...

Page 19: ...ease of reading on the displays Click Write on the bottom of the Name panel to send the user names to the Host Head If the Write and Reset buttons are greyed out first click Read to enable them type...

Page 20: ...d Troubleshooting For maintenance and troubleshooting please see the POLI User Guide 8 mSquad Technical Specifications All specifications and listed sensors are subject to change without notice Please...

Page 21: ...mSquad User s Guide 21 Technical Support and mPower Contacts mPower Electronics Inc 3046 Scott Blvd Santa Clara CA 95054 Phone 408 320 1266 Fax 669 342 7077 info mpowerinc com www mpowerinc com...

Reviews: