VOXI User’s Guide
2
Read Before Operating
This manual must be carefully read by all individuals who have or will have the responsibility of
using, maintaining or servicing this product. The product will perform as designed only if it is
used, maintained and serviced in accordance with the manufacturer’s instructions. The user
should understand how to set the correct parameters and interpret the obtained results.
WARNINGS !
•
Never operate the monitor when the cover is removed.
•
Remove the monitor cover only in an area known to be non-hazardous.
•
Use only mPower’s sensor and accessories. Substitution of components will impair
suitability for intrinsic safety and void warranty.
•
The instrument should be calibrated after installation before initial use and checked by
exposing it to a known concentration calibration on a regular basis.
•
Ensure that the gas inlet is not blocked.
•
Make sure that all filters are clean and replaced on a regular basis.
•
Remove the sensor only if necessary for repair. Zero and span calibration are required once
the sensor is moved.
Special Conditions for Safe Use
The VOXI monitor must be calibrated if it does not pass a bump test, when a new sensor has
been installed, or at least once every 180 days, depending on use and sensor exposure to poisons
and contaminants.