VOXI User’s Guide
20
6.2.3
Sensor Cleaning
1.
Unscrew the sensor cover and pull the sensor straight out, using a
slight rocking motion if necessary.
2.
Put the sensor into a beaker and cover it with pure methanol or
ethanol.
3.
Put the beaker into an ultrasonic cleaning bath and sonicate
for 15 minutes. Take out sensor and dry it, preferably using a
gentle stream of clean air to blow the residual liquid out of
the sensor.
4.
Always re-calibrate the PID after cleaning the sensor.
6.3 Replacing Pump
For replacement of pump,
please contact an authorized mPower service center.