Moxy Overview
1.
On/Off Button. Hold for 3 seconds to turn the sensor on or off. Holding
the button for more than 10 seconds will cause a hard reset of the
sensor microcontroller. A hard reset will not delete any stored data, but
it will reset the clock that is used for time stamps in the stored data.
Short click the button to make the green LED indicate the battery
charge level. A short click also marks a lap in the data stored on the
sensor.
2.
Yellow Indicator LED.
On solid indicates connection to the Moxy PC App.
A blink after a short button click indicates that the sensor is on.
3.
USB Port. This is a micro USB port. It is used for charging the Sensor or
downloading data.
4.
Green Indicator LED. This is used to indicate battery status.
Short blinks while the USB cable is plugged in indicate charging in
progress.
After a short button click, the number of blinks indicates charge
level
5.
The ANT+ antenna is located here.
6.
Pressure Equalization Vent Port. The port is covered with a waterproof
membrane on the inside.
7.
Serial Number. This serial number is used in ANT+ to identify this
individual Moxy sensor. It is also stored in data files to identify which
Sensor acquired the data.
8.
Emitter Optode
9.
Short-Spacing Optode
10.
Long-Spacing Optode
5