C A L I B R A T I O N S E T - U P
C o n t i n u e s …
DECIMAL POINT
Prompts to decimal point selection
Scroll Right
to change decimal point location
Enter
to validate
LOW CAPACITY
This value is the transducer’s minimum capacity
provided by the manufacturer.
Scroll Up
to change the number
Scroll Right
to change digit
Enter
to validate
RATED OUTPUT
The TorqueLab
â
can accept a rated output ranging
from 1.700 to 2.300 mV/V and 5 V (Brushless
transducers). This value is provided by the
manufacturer and needs to be known if performing
a Fast Calibration. If you intend to do a True
Calibration you can enter 2.000 mV/V
Scroll Up
to change the number
Scroll Right
to change digit
Enter
to validate
FAST CALIBRATION
This method of calibration is based on a two-point
(zero and full scale) calibration system, which
doesn’t compensate for transducers nonlinearly.
It’s a lesser accurate method than the True calibration.
Scroll Up
to change the method
Enter
to validate
Manual
Clear
+
B
Data
Torque
mV.V
+
B
Auto
Clear
Manual
Clear
Downloa
Set-up
+
B
Data
Torque
Threshold
+
Torque
Threshold
B
B
Auto
Clear
Downloa
+
B
Data
+
Manual
Clear
Set-up
Note:
The decimal point is now at a fixed
location, previously determined during
the HIGH CAPACITY set-up.
Note:
When choosing this method, the
calibration program will be completed and
the display will return to the operating
mode.