Page 8
The EZ-TorQ Torque Tester offers three different modes of displaying torque information during operation
(Track, Peak and First Peak). The user will determine which mode is best suited for the application. The use of the three
modes is described below.
T R A C K
This mode constantly tracks increasing or decreasing torque
variations. Use this mode to monitor varying torque on motors
and machinery. Also for calibration and testing of dial torque
products (small wrenches or dial screwdrivers).
P E A K
The display retains the highest torque applied. Use this mode
during calibration or testing of any hand type torque wrench
(dial, beam, and screwdriver), as well as power tools.
F I R S T P E A K
The display holds the first detected torque peak applied.
Before any torque is applied, the display show dashes in the
torque value area. Once peak is detected, the display will
show the torque value. If a second peak is detected then it
will be displayed in the lower right (in small reverse video).
The function of this mode is primarily used for testing and
calibrating click type mechanical torque wrenches. The
tester captures the point where the wrench first clicks. As
additional force is applied a secondary torque is displayed.
This mode may be used for operator training on the correct
use of operating a click wrench. Always apply torque smoothly
to avoid false first peak readings.
M o d e S e l e c t i o n
Track
+12.010
Lbf.in
Menu
Peak
+22.350
ozf.in
Menu
S e t u p S e l e c t i o n s
This function allows the user to turn on the display backlight for low light
environments.
Use the
Up / Down scroll keys
to select the desired setting and press
Enter
to select the setting.
Backlight
BACKLIGHT SETTING
Backlight: OFF
FPK
500 Hz
+23.615
lbf.in
Menu
+24.755
Corporate Headquarters:
1080 North 11th Street, San Jose, CA 95112 Phone: (408) 292-2214 Fax: (408) 292-2733
www.mountztorque.com