MOTUS GI PURE-VU SYSTEM INSTRUCTIONS FOR USE
9.0 TECHNICAL SPECIFICATIONS
9.1. Environmental Standards
The table below describes the recommended environmental standards required for correct operation of
Motus Pure-Vu System.
SPECIFICATIONS
Operating Temperature
50° to 86° F
10° to 30° C
Humidity
10% to 80%
(Non-condensing)
Altitude
Up to 81 Kpa
9.2. Shipping and Storage
Recommended Shipping and Storage Temperature
SPECIFICATIONS
Shipping
Storage
14° to 140° F / -10° to 60° C
59° to 86° F / 15° to 30° C
9.4. Mechanical Properties and Sizes
Following are the mechanical properties and size for the Pure-Vu Workstation and Loading Fixture:
Workstation
Loading Fixture
Workstation Outer Dimensions
250mm x 476mm x 358mm
[10in x 19in x 14in]
584mm x 267mm x 241mm
[23in x 10.5in x 9.5in]
Workstation Weight
18kg (40lbs)
7.5kg (15.5lbs)
Power Supply
100 - 240 VAC / 50 - 60 Hz
100 - 240 VAC / 50 - 60 Hz
9.3. Performance Limitations
MEDIA
MAX PRESSURE
FLOW
Cleansing
Air
1.55 Bar [23PSI]
Up to 1350 cc/min
Water
3.5 Bar [52PSI]
Up to 645 cc/min
Pressure
Vacuum
Up to (-) 0.5 Bar
NA
Internal sensor is
outside optimal
operating range.
(System beeps
twice and “Status”
light blinks orange.)
• Verify system is outside optimal operating range by listening for audible indicators and confirming
visual indicator. (System beeps twice and “Status” light blinks orange.)
• Be aware that the auto-purge function may not be available while system is outside of optimal
operating range. Manual purge should be used until system returns to optimal operating range.
• Check tubing for obstructions or bending and mitigate as needed.
• Verify system return
s
to optimal operating range by visually confirming “Status” light has returned
to
solid green.
• No other action is required of the operator.
If irrigation is
outside optimal
operating range.
(System beeps
twice and “Status”
light blinks orange
and “Low” light
blinks green.)
• Verify system is outside optimal operating range by listening for audible indicators and confirming
visual indicators. (System beeps twice and “Status” light blinks orange and “Low” light blinks green.)
• Check tubing for obstructions or bending and mitigate as needed.
• Be aware that while the system is outside of optimal operating range system will automatically shift
to or continue to run only in Low mode. Medium and High cleansing modes are deactivated until the
system is restarted or drain operation is completed.
• No other action is required of the operator.
*If the above does not remedy the warning indication, contact a Motus GI representative.