SYS TEST
LOW MED HIGH
ERROR
STATUS
DRAIN
A
B
C
D
E
F
G
Open Pump Head
E
A
B
D
C
13
MOTUS GI PURE-VU SYSTEM INSTRUCTIONS FOR USE
FIGURE 8:
Disconnect the air line from the inflation hub (A). Turn OFF the inflation control box power (B). Move the
locking ring (C) to the distal end (unlocked position) and remove the sealing plug (D). Do not dispose of the sealing
plug, as it will be used in the unloading process. Inspect the distal end of the Pure-Vu Oversleeve/colonoscope interface,
making sure the colonoscope is flush with the upper lip of the Pure-Vu Oversleeve as shown (E). Remove the Pure-Vu
Oversleeve with colonoscope from the Loading Fixture. The Pure-Vu Oversleeve and colonoscope are now ready for
assembly to the Pure-Vu Umbilical Section, Workstation controller, colonoscope light source and video processor.
FIGURE 9:
Connect the main connector (A) to the main connector socket in the Workstation (B). Place the irrigation
tubing (C) into the irrigation pump marked in blue (D) between marked blue lines, then close the irrigation pump
head. Remove the irrigation tube spike’s cap (G), then connect the irrigation tubing to the irrigation bag (E) and
ensure that the irrigation line clamp (F) is open.
4.2.2. Oversleeve Assembly
(cont...)
4.2.3. Attaching the Umbilical Section to the Workstation
The Workstation should be cleaned prior to attaching the Umbilical Section to the Workstation by
thoroughly wiping down the Workstation with a viricidal cloth (Opti-cide, Cavicide or similar surface
disinfectant of choice) that has an active ingredient of isopropyl alcohol (17% - 21%).
Remove the Umbilical Section from the tray (item #2 in the tray). To attach the Umbilical Section to the
Workstation complete the following steps: