background image

3.  Place the microscopic specimen slide on the stage, making sure that the cover slip is raised. 

Lift the specimen holder clips (5) pressing downwards on the lower part, and slide the 
specimen slide under the clips. Release pressure so that the slide is firmly held in place.   

 

4.  Make sure that the sample on the slide is in the optical path.  

 

5.  While looking through the eyepiece (1), turn the coarse focus knob until the sample is in 

focus. 

 

6.  Readjust the focus with the fine focus knob (13) until the image appears sharp. 

 

C.  Adjusting the aperture of the diaphragm. 

 

The diaphragm (8) should not be used to regulate the light intensity. Its function is to obtain a 
high resolution of the specimen and to provide contrast in the image. Smaller apertures will 
deliver higher contrast to image, although closing the aperture too much will reduce 
resolution. The best way to obtain the correct aperture of the diaphragm is to experiment. 
Suggested apertures for each objective are as follows: 
 
  

OBJECTIVE 

 

APERTURE OF IRIS 

 
  

4X 

 

 

From fully closed to 1/8 open. 

  

10X 

 

 

From 1/8 to 1/4 

  

40X 

 

 

From 1/4 to 1/2 

  

100X (optional)  

From 1/2 to 3/4 

 

D.  Changing magnification. 

 

1.  Position the objective 10X (4) in the optical path. 

 

2.  This microscope has already been parfocalised, although it is possible that small differences 

exist between the objectives. It may then be necessary to adjust the focus slightly with the 
fine focus knob (13). 

 

3.  When changing to the 40X and 100X objectives (optional), it must be done with great care, 

ensuring that the objectives do not cause damage to the front lens. 

 

4.  In order to obtain maximum resolution of the 100X (optional), it is necessary to apply 

immersion oil between the cover slip of the slide and the front lens of the objective.   

 

a.  Only a very small amount of immersion oil is needed, a drop should be enough. 

 

b.  If air bubbles appear they can be removed by gently rotating the nosepiece back 

and forth.  

 

c.  After viewing, all parts that have come into contact with the oil must be cleaned 

using a soft cotton cloth, lightly dampened with Xylene. If the 100X objective is not 
cleaned, the oil will dry, and it will not be possible to see through it; permanent 
damage could also occur.  

 

NB. Immersion oil must ONLY be used with the 100X objective, as it is the only one 
specially prepared for it. If any other objective comes into contact with the oil, it must 
be cleaned immediately.  

 
 

 
 
 
 
 

-5- 

 

Summary of Contents for F-11 LED Series

Page 1: ...Instruction Manual F 11 LED Series Copyright 02 02 Motic Microscopes European Division...

Page 2: ...volving nosepiece 4 Objectives 6 Stage 7 Condenser 5 Specimen slide holder clips 8 Diaphragm 9 Filter holder 10 Head 11 Arm 12 Coarse focus knob 13 Fine focus knob 14 Illuminator 2 Eyepiece tube 15 Ba...

Page 3: ...15 model with additional notes applying specifically to other models in the series Unpacking All components of the microscope have been carefully packed to ensure they reach you in perfect condition W...

Page 4: ...mination for enhanced resolution and image contrast 7 Focusing Knobs 12 13 Situated on both sides of the arm of the microscope the larger or coarse focusing knob initially brings the specimen into foc...

Page 5: ...tional From 1 2 to 3 4 D Changing magnification 1 Position the objective 10X 4 in the optical path 2 This microscope has already been parfocalised although it is possible that small differences exist...

Page 6: ...ece tube between the eyepiece and the diopter adjustment collar With a watchmakers screwdriver loosen the screw until the eyepiece can be removed without removing the screw 2 Do not loosen any of the...

Page 7: ...or the illuminator d With carefully take out of the bulb and pull outwards to disconnect it from the socket e Do not touch the new bulb with bare hands Use a clean cloth to insert the bulb pins into t...

Page 8: ...the upward movement of the stage The rack stop screw comes pre adjusted by the factory for standard slides with a 0 17mm thick cover slip However for observing other types of slides adjustment may be...

Page 9: ...side down Wrong cover slip used with slide Light too bright Dirty condenser Clean objective Clean eyepiece Replace the slide with the cover slip facing upwards Use 0 17mm thick cover slips Reduce ligh...

Page 10: ...n Warrantee All MOTIC microscopes are warranted against any manufacturing defect for a 5 year period Damage occurring by any unauthorised repair work or occurring through misuse or modification of the...

Reviews: