E.  Adjusting diopter (only for model F-1125). 

 
This model has a dual head, i.e. with two head 
holder tubes: one inclined, and the other vertical. 
This permits use by two people at the same time. 

 

1.  To use the inclined eyepiece tube, focus on 

sample using the coarse (12) and fine focus 
knobs (13). 

 

2.  To compensate for the difference in vision of 

the two observers, the user of the vertical 
eyepiece tube can adjust focus by turning the 
diopter adjustment collar. (Fig. 2). 

 
 

F.  Adapting a camera (only for model F-1125). 

 

These are general guidelines on how to install a photographic or video camera to the vertical 
eyepiece tube. You should know, however, that each model of camera has its own specific 
instructions, and may be installed in a distinct manner.  
There exist different types of adapters to install a camera on a microscope. In some of them the 
eyepiece must be removed, whilst for others it is not necessary.  

 

1.  To remove the eyepiece, find the tiny locking screw (Fig. 3) on the eyepiece tube, between 

the eyepiece and the diopter adjustment collar. With a watchmakers screwdriver, loosen the 
screw until the eyepiece can be removed, without removing the screw.   

 

2.  Do not loosen any of the three screws on the diopter adjustment collar.  

 

Maintenance    

 

WARNING:

  FOR YOUR OWN SAFETY SWITCH OFF AND DISCONNECT THE MICROSCOPE FROM 

ANY ELECTRICAL SOURCE BEFORE ATTEMPTING ANY MAINTENANCE PROCEDURE IN ORDER TO 
AVOID THE RISK OF ELECTROCUTION.  
 
IF THE MICROSCOPE REQUIRES ANY MAINTENANCE OR REPAIR NOT APPEARING IN THIS 
MANUAL, CONSULT YOUR DISTRIBUTOR.  
 
 

A.  Optical maintenance. 

 

Do not attempt to disassemble any optical component.  
Prior to cleaning any of the lens  surfaces, remove dust particles using a fine brush, specifically for 
cleaning lenses. Alternatively, use low pressure compressed air, available in shops selling 
photography equipment. 

 

1.  Cleaning the eyepiece. 

 

a.  Do not remove eyepiece from eyepiece tube. 
 
b.  Clean only the outer surface, misting the lens with breath.  
 
c.  Dry by wiping with lens paper in circular movements, from centre, outwards. Do not wipe 

lenses when dry, as they can easily be scratched. 

 
 
 
 
 

-6- 

Fig. 2 

Eyepiece 

locking 

screw 

Diopter 

adjustment 

collar 

Summary of Contents for F-11 LED Series

Page 1: ...Instruction Manual F 11 LED Series Copyright 02 02 Motic Microscopes European Division...

Page 2: ...volving nosepiece 4 Objectives 6 Stage 7 Condenser 5 Specimen slide holder clips 8 Diaphragm 9 Filter holder 10 Head 11 Arm 12 Coarse focus knob 13 Fine focus knob 14 Illuminator 2 Eyepiece tube 15 Ba...

Page 3: ...15 model with additional notes applying specifically to other models in the series Unpacking All components of the microscope have been carefully packed to ensure they reach you in perfect condition W...

Page 4: ...mination for enhanced resolution and image contrast 7 Focusing Knobs 12 13 Situated on both sides of the arm of the microscope the larger or coarse focusing knob initially brings the specimen into foc...

Page 5: ...tional From 1 2 to 3 4 D Changing magnification 1 Position the objective 10X 4 in the optical path 2 This microscope has already been parfocalised although it is possible that small differences exist...

Page 6: ...ece tube between the eyepiece and the diopter adjustment collar With a watchmakers screwdriver loosen the screw until the eyepiece can be removed without removing the screw 2 Do not loosen any of the...

Page 7: ...or the illuminator d With carefully take out of the bulb and pull outwards to disconnect it from the socket e Do not touch the new bulb with bare hands Use a clean cloth to insert the bulb pins into t...

Page 8: ...the upward movement of the stage The rack stop screw comes pre adjusted by the factory for standard slides with a 0 17mm thick cover slip However for observing other types of slides adjustment may be...

Page 9: ...side down Wrong cover slip used with slide Light too bright Dirty condenser Clean objective Clean eyepiece Replace the slide with the cover slip facing upwards Use 0 17mm thick cover slips Reduce ligh...

Page 10: ...n Warrantee All MOTIC microscopes are warranted against any manufacturing defect for a 5 year period Damage occurring by any unauthorised repair work or occurring through misuse or modification of the...

Reviews: