E. Adjusting diopter (only for model F-1125).
This model has a dual head, i.e. with two head
holder tubes: one inclined, and the other vertical.
This permits use by two people at the same time.
1. To use the inclined eyepiece tube, focus on
sample using the coarse (12) and fine focus
knobs (13).
2. To compensate for the difference in vision of
the two observers, the user of the vertical
eyepiece tube can adjust focus by turning the
diopter adjustment collar. (Fig. 2).
F. Adapting a camera (only for model F-1125).
These are general guidelines on how to install a photographic or video camera to the vertical
eyepiece tube. You should know, however, that each model of camera has its own specific
instructions, and may be installed in a distinct manner.
There exist different types of adapters to install a camera on a microscope. In some of them the
eyepiece must be removed, whilst for others it is not necessary.
1. To remove the eyepiece, find the tiny locking screw (Fig. 3) on the eyepiece tube, between
the eyepiece and the diopter adjustment collar. With a watchmakers screwdriver, loosen the
screw until the eyepiece can be removed, without removing the screw.
2. Do not loosen any of the three screws on the diopter adjustment collar.
Maintenance
WARNING:
FOR YOUR OWN SAFETY SWITCH OFF AND DISCONNECT THE MICROSCOPE FROM
ANY ELECTRICAL SOURCE BEFORE ATTEMPTING ANY MAINTENANCE PROCEDURE IN ORDER TO
AVOID THE RISK OF ELECTROCUTION.
IF THE MICROSCOPE REQUIRES ANY MAINTENANCE OR REPAIR NOT APPEARING IN THIS
MANUAL, CONSULT YOUR DISTRIBUTOR.
A. Optical maintenance.
Do not attempt to disassemble any optical component.
Prior to cleaning any of the lens surfaces, remove dust particles using a fine brush, specifically for
cleaning lenses. Alternatively, use low pressure compressed air, available in shops selling
photography equipment.
1. Cleaning the eyepiece.
a. Do not remove eyepiece from eyepiece tube.
b. Clean only the outer surface, misting the lens with breath.
c. Dry by wiping with lens paper in circular movements, from centre, outwards. Do not wipe
lenses when dry, as they can easily be scratched.