background image

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 

F-1115 

 
 
 
 
 
 

-2- 

1. Eyepiece 

3. Revolving 
    nosepiece 

4. Objectives 

6. Stage 

7. Condenser 

5. Specimen slide 
    holder clips 

8. Diaphragm 

9. Filter holder 

10. Head 

11. Arm 

12. Coarse focus 

knob 

13. Fine focus 

knob 

14. 

I

lluminator 

2. Eyepiece tube 

15. Base 

16. Intensity 

control 

Summary of Contents for F-11 LED Series

Page 1: ...Instruction Manual F 11 LED Series Copyright 02 02 Motic Microscopes European Division...

Page 2: ...volving nosepiece 4 Objectives 6 Stage 7 Condenser 5 Specimen slide holder clips 8 Diaphragm 9 Filter holder 10 Head 11 Arm 12 Coarse focus knob 13 Fine focus knob 14 Illuminator 2 Eyepiece tube 15 Ba...

Page 3: ...15 model with additional notes applying specifically to other models in the series Unpacking All components of the microscope have been carefully packed to ensure they reach you in perfect condition W...

Page 4: ...mination for enhanced resolution and image contrast 7 Focusing Knobs 12 13 Situated on both sides of the arm of the microscope the larger or coarse focusing knob initially brings the specimen into foc...

Page 5: ...tional From 1 2 to 3 4 D Changing magnification 1 Position the objective 10X 4 in the optical path 2 This microscope has already been parfocalised although it is possible that small differences exist...

Page 6: ...ece tube between the eyepiece and the diopter adjustment collar With a watchmakers screwdriver loosen the screw until the eyepiece can be removed without removing the screw 2 Do not loosen any of the...

Page 7: ...or the illuminator d With carefully take out of the bulb and pull outwards to disconnect it from the socket e Do not touch the new bulb with bare hands Use a clean cloth to insert the bulb pins into t...

Page 8: ...the upward movement of the stage The rack stop screw comes pre adjusted by the factory for standard slides with a 0 17mm thick cover slip However for observing other types of slides adjustment may be...

Page 9: ...side down Wrong cover slip used with slide Light too bright Dirty condenser Clean objective Clean eyepiece Replace the slide with the cover slip facing upwards Use 0 17mm thick cover slips Reduce ligh...

Page 10: ...n Warrantee All MOTIC microscopes are warranted against any manufacturing defect for a 5 year period Damage occurring by any unauthorised repair work or occurring through misuse or modification of the...

Reviews: