GENERAL SAFETY INSTRUCTIONS

Please be sure to read these instructions before using the biological microscope. Additional 

information is available upon request from our maintenance department or authorized 

agency. To ensure safe operation and guarantee good performance of the microscope please 

pay attention to the precautions and warnings specified in the Operation Instructions.

INSTRUMENT SAFETY

The SILVER250 Series biological microscope has been designed, manufactured and 

inspected according to the EN 61010-1:2001 Safety Requirements for Electrical Equipment 

for Measurement, Control and Laboratory Use.

UNPACKING, TRANSPORTATION & STORAGE

The original shipping container, a foam box in a fiberboard carton, should be kept for use in 

long term storage or return shipment. When unpacking, please check the components 

according to the packing list. Please comply with the temperature requirements for transport 

and storage specified in the appendix of this manual. Set up, use and store the unpacked 

microscope on a firm and flat workbench. Please do not touch the optical lens surfaces.

WASTE DISPOSAL

Important:

 Any damaged biological microscope must not be treated as general waste; it 

should be disposed of according to relevant regulations.

OPERATION

When using the biological microscope, please pay attention to the following safety instruc-

tions:

• If it is used for any purpose other than the specified ones, including any individual compo-

nent or part, the manufacturer will not take any responsibility.

• After-sales service or repair done by unauthorized personnel will void the warranty.

• Anyone who uses the instrument should receive instruction on the proper handling of the 

instrument and safety practices for microscopy. The biological microscope shall be placed 

only on a firm, flat workbench for operation.

• Since the biological microscope is a precision instrument, improper operation will impair or 

spoil its performance.

• The power unit is integrated in the main unit of the biological microscope: the grid supply 

voltage is within 100-240V~50Hz.

The biological microscope must be connected only to the normal power socket with a 

grounding terminal. Any extension cord without ground protection is not allowed to avoid 

failure of the protection function.

SILVER

250

 

SERIES

 

|  02 

  

SAFETY INSTRUCTIONS

SILVER

 LINE

Summary of Contents for 1100104800031

Page 1: ...SILVER250 SERIES INSTRUCTION MANUAL ENGLISH SILVERLINE...

Page 2: ...uch the optical lens surfaces WASTE DISPOSAL Important Any damaged biological microscope must not be treated as general waste it should be disposed of according to relevant regulations OPERATION When...

Page 3: ...ive latent infective toxic radioactive or other hazardous samples Therefore when examin ing any such sample you must comply with the relevant laws and regulations in particular the provisions related...

Page 4: ...SERIES 04 SILVER250 SERIES The SILVER250 Series stands out for its innovation quality and student proof features offering a wide range of microscopes for educational purposes in high schools up to adv...

Page 5: ...he minimum position before turning on or off the power switch 4 Rotate the brightness control to the desired illumination 5 After use turn the brightness control to the minimum position and then turn...

Page 6: ...itting properties as glass Using oil slight ly increases the resolution and brightness of the image Usually a very thin slide size 1 is used for oil immersion because at this magnification the working...

Page 7: ...ll the correct driver IMPORTANT The next time you connect the SILVER250 Series microscope to a computer make sure the USB cable is connected to the same USB port that was used during initial installat...

Page 8: ...for 0 17 transmitted light objective No immersion oil or non specified immersion oil for 100X oil immersion objective Aperture iris diaphragm is not set accurately Incorrect opening of aperture diaphr...

Page 9: ...access the LED The LED is integrated on to a circuit board This LED circuit board is held into the illumina tor housing by 2 screws Remove the 2 screws from the illuminator housing to remove the LED c...

Page 10: ...tube 30 20 Length of mechanical tube Maximum field number Hinge type interpupillary distance adjustment range Observation angle Infinite 20mm 48 to 75mm 30 Triocular tube 30 20 Length of mechanical t...

Page 11: ...epair by unauthorized parties accident alteration shipping misuse or abuse is not covered Warranty service is provided by Motic authorized technicians Determination of warranty is at the technician s...

Page 12: ...52 2882 2792 Motic Europe S L U Spain Pol gon Industrial Les Corts C Les Corts 12 08349 Cabrera de Mar Barcelona Spain Tel 34 93 756 6286 Fax 34 93 756 6287 CCIS is a trademark of Motic Incorporation...

Page 13: ...he appropriate immersion oil to use IMPORTANT The working distances of the 40XRD and 100XRD objectives to the slide surface are very small and although these objectives are sealed to prevent oil conta...

Page 14: ...f the glass The finish of the microscope is hard epoxy and is resistant to acids and reagents Clean this surface with a damp cloth and mild detergent Periodically the microscope should be disassembled...

Page 15: ...view Is the Abbe condenser in the correct position Dust or hairs seem to be moving in the image The iris diaphragm is not open wide enough Slowly open the diaphragm to increase the size of the openin...

Reviews: