Page 7
Operation 12. 21. ’04
Charging Battery
The battery is built into Low Speed Handpiece Module.
NOTE
•
The battery is not charged when the unit is shipped
from the factory and must be charged before using
the unit.
•
If the plug for the AC adapter does not fit the socket,
it is the user’s responsibility to find a suitable plug
adapter.
1.
Line up the arrow ( ) on the AC adapter’s connector with the triangle (
◀
) above its jack on
the left side of Low Speed Handpiece Module and plug it in. Then plug the adapter into the
electrical power receptacle.
WARNING
•
Do not use the unit when the AC adapter is connected.
2.
The Amber Charge LED on the back of Low Speed
Handpiece Module starts flashing on and off and then,
after a few seconds, it will stop flashing and stay on to
show that the battery is being charged. It takes
about 60 minutes to fully charge the battery.
3. Amber Charge LED goes out when the battery is fully charged.
4. Disconnect the AC adapter from Low Speed Handpiece Module and unplug it.
WARNING
•
Never operate the unit with an external power supply.
•
If an electrical storm occurs while the battery is being charged, do not touch the AC adapter or
the charger’s power supply cord as there would be a risk receiving an electric shock.
NOTE
•
Do not pull or yank the cord when disconnecting the AC adapter.
Amber Charge LED On
Low Speed Handpiece Module
AC Adapter Plug