Operation 2018-09-21
33
10.
Electromagnetic Disturbances (EMD)
The Lubrina (Model:HIM-1, Type:CE and EX230V, hereafter "this device") conforms to IEC 60601-1-2:2014 Ed. 4.0, the relevant
international standard for electromagnetic disturbances (EMD).
The following is the “Guidance and Manufacturer's Declaration” which is required by IEC 60601-1-2:2014 Ed. 4.0, the relevant
international standard for electromagnetic disturbances.
This is a Group 1, Class B product according to EN 55011 (CISPR 11).
This means that this device does not generate and/or use internationally radio-frequency energy, in the form of electromagnetic
radiation, inductive and/or capacitive coupling, for the treatment of material or inspection/analysis purpose and that it is suitable for
use in domestic establishments and in establishments directly connected to a low voltage power supply network which supplies
buildings use for domestic purposes.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
This device is intended for use in the electromagnetic environment specified below.
The customer or the user of this device should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment – Guidance
Conducted disturbance
CISPR 11
Group 1
Class B
This device uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
Radiated disturbance
CISPR 11
Group 1
Class B
This device is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonic current
*1
IEC 61000-3-2
Class A
Voltage fluctuations and flicker
IEC 61000-3-3
Clause 5
*1: Although this device is not applicable to Harmonics test since the rated power is less than 75 W, it has been tested as a reference
according to limits for Class A.
•
The use environment of this device is the Professional healthcare facility environment.
•
This device needs special precautions regarding EMD and needs to be installed and put into service according to
the EMD information provided in the ACCOMPANYING DOCUMENTS.
•
Use of parts other than those accompanied or specified by J. MORITA MFG. CORP. could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this device and result in improper operation.
•
Do not use this device as adjacent or stacked as possible with other.
When adjoining or stacking is necessary, use it after observing whether this equipment and other equipment work
properly.
•
Portable and mobile RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm to any part of the HIM-1, including cables specified by the
manufacturer.