Operation 2018-05-21
27
Spray Can Replacement
Replace a can if its empty lamp starts to blink.
(See page 13 “(6) Spray Cans”.)
Empty Drainage from Air Regulator
*
Empty the drainage from the air regulator about once a
month.
1.
Pull up the dial on the air regulator and turn it until the
pressure drops to 0 MPa.
If the drain valve is opened without releasing the pressure inside
the regulator, the water in the drain could spray out into your
eyes.
2.
Put a container under the regulator and then press the drain
valve.
The dirty water in the regulator will drain out.
If water is inadvertently mixed with the air, handpieces could be
severely damaged.
Empty Lamps
Drain Valve
Air Regulator Dial
Air Regulator