29
Instructions for Use 2019-01-16
8.1 Daily Cleaning
1. Remove the oil absorbent sheet, wipe off any remaining oil, and
wipe off the unit with ethanol (70 vol% to 80 vol%).
p. 19
Clean the couplings with ethanol (70 vol% to 80 vol%).
Be sure to check that the O-rings are in place and not damaged.
If you find any damaged O-rings, replace them with new ones immedi
-
ately.
p. 32
8.1.1 Main Unit
8.1.2 Couplings
8 Cleaning the Device and Replacing Parts
• To prevent infection, always wear appropriate personal protective equipment such as surgical gloves, when using or cleaning the device.
•
To prevent malfunction, be sure to turn the Power Switch off before performing the Lubrina 2 maintenance.
2. Remove the door oil absorbent sheet and front door sheet stopper,
and clean the inside of the front door in the same way as the unit
surface.
p. 20
“6.6 Door Oil Absorbent Sheet and Front Door Sheet Stopper”
* Order parts from your local dealer or J. MORITA OFFICE.
* Reinstall all parts after cleaning.
Do not use disinfectants that contain chlorine-based cleaners or benzalkonium chlorinated
aromatics. These solutions could discolor or damage plastic and corrode metal.
Be sure to use a soft cloth, such as gauze, to wipe off the unit. Otherwise, the unit surface
will get scratched.
Note that lack of cleaning will result in oil overflowing and contaminating the installation
area.
* If there is dirt that is hard to remove, use a soft cloth with water and a little neutral deter-
gent.
* Be sure not to lose the O-rings.
Do not autoclave the couplings. Doing so could damage the couplings.
Never attach a handpiece to a coupling if the O-ring is not properly in place. Doing so
could result in improper lubrication, handpiece damage, or severing the O-ring, so that the
handpiece becomes stuck and difficult to detach.
Use a clean gauze to clean couplings. Debris inside the connection could cause the
handpiece to malfunction.
Front Door
Right
Wrong