27
7.4 Other Maintenance Modes
Instructions for Use 2019-01-16
Hold Down
7.4 Other Maintenance Modes
7.4.2 Flushing Mode
Use Flushing Mode as needed, such as when black colored oil is discharged from the handpiece.
Flushing Mode will set a longer time for lubrication and blowing air to boost the effectiveness of the flushing ef
-
fect. Each handpiece requires approximately two minutes to complete the process.
* The term “flushing”, as used in here shall mean ejecting and removing cutting debris and foreign objects from inside the handpiece.
2. Hold down the Line Selection Switch for the line where the handpiece is attached.
Air Blow Mode starts after approximately two seconds. You can release the switch when the process starts.
The process automatically stops after 30 seconds.
You can stop the process manually by pressing the Start/Stop Switch.
* Air Blow Mode can be set and performed on only one line at a time.
1. Attach the handpiece and close the front door.
7.4.1 Air Blow Mode
█
Procedures
Use Air Blow Mode to remove excess oil when performing lubrication for handpieces other than J. MORITA prod-
ucts or after direct lubrication from a spray can.
• Always wear appropriate personal protective equipment such as surgical gloves, when using or cleaning the device.
• A large amount of oil mist is produced during Flushing Mode. Be sure to use Flushing Mode in a well-ventilated location.