34
ENGLISH
Rated maximum
output power
of transmitter W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
d =
√
P
80 MHz to 800 MHz
d = 0.6
√
P
800 MHz to 2.7 GHz
d = 1.2
√
P
0.01
0.1
0.06
0.12
0.1
0.31
0.19
0.38
0.1
0.1
0.6
1.2
10
3.1
1.9
3.8
100
10
6
12
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflec-
tion from structures, objects, and people.
20. TECHNICAL FEATURES
20.1 Pump unit
Features
Specification
Power Source
DC12V, 1250 mA by Adaptor
Power consumption
12V DC, 1150 mA
Device Dimension
(L x W x H)
25 x 12,2 x 5,7 cm
Device Weight
1.0 kg or 2.2lb (battery and adaptor included)
Environment
Atmospheric Pressure
700 hPa to 1013,25 hPa
Environment
Temperature
Operation: 5°C to 40°C (41°F to 104°F)
Storage: -10°C to 45°C (14°F to 113°F)
Shipping: -10°C to 45°C (14°F to 113°F)
Environment
Humidity
Operation: 30% to 75% non-condensing
Storage: 10% to 75% non-condensing
Shipping: 10% to 90% non-condensing
Classification
Type BF, Class II, IP2X; IP22 (with carrying bag)
Applied part: Air Seat Cushion
Not suitable for use in the presence of a flammable anaesthetic
mixture (No AP or APG protection)
Pressure Range
Five selectable settings
Cycle Time
10, 15, 20 minutes
Battery Pack
3.78V Li-ion Battery x2 (2.95Ah typical)
Battery Operation
Time
Approx. 12 hours
Battery Recharge
Time
Approx. 4.5 hours
Specification