13
ENGLISH
CAUTION!
If there is a possibility of electro-magnetic interference with mobile phones, please
increase the distance (3,3m) between devices or turn off the mobile phone.
13. OPERATION
NOTE: Always read the operating instruction before use
13.1 General operation
1. Switch on the main power switch found on
the side of the pump.
2. Once the pump begins delivering air into the
mattress it will take approximately 30 minutes
to fully inflate the mattress.
3. Every time when mattress is first setup
for use, It will be forced to execute Maxfirm
for the quickest inflation. The low-pressure
indicator (yellow LED) will light up when the
mattress is not fully inflated at initial inflation.
If the appropriate pressure is reached, the low-
pressure indicator (yellow LED) will go off.
4. (LAD680) When the initial inflation (Max
firm process) is completed, the system will
enter the “alternating mode”.
5. (LAD680) According to the weight and
height of the patient, adjust the pressure
setting to the most comfortable level without
bottoming out, then the pressure in mattress
will slowly increase to the intended value after
the air mattress is ready to use.
4. (LAD690) When the initial inflation
(Maxfirm process) is completed, the system will
automatically enter the Auto-Adjust process.
5. (LAD690) According to the weight and
height of the patient, the system will automati-
cally adjust an appropriate pressure setting to
ensure the patient’s comfort, and then enter to
alternating mode.