Replace pump if damaged
Internal float switch stuck or
bad
Un-stick or replace internal float switch
Green Button is bad
Replace green button
Icon on touch screen says “!”
with “Formalin Low” at bottom
of screen
Internal Formalin storage
Tank is low
Add formalin to internal tank using transfer pump
Leak Icon (sensor #1, inside
station, beneath formalin storage
tank) showing on screen
There is a formalin leak or
sensor is faulty
Repair leak or replace faulty sensor
FORMALIN COLLECTION SYSTEM
Formalin collection “!” icon
showing on touch screen
Formalin collection
Jug/Carboy is full
Empty formalin Jug/Carboy
Float switch disconnected
Plug float switch into station
Check wires in connector are not pulled out
Float switch “bulb” is stuck in
the “full” (up) position
Unscrew cap on Jug/Carboy and make sure metal
“bulb” on float switch can move up and down the
rod without getting stuck. Or replace float switch.
LEAK “!” Icon showing on touch
screen (sensor #2, located in
formalin collection cart)
Sensor disconnected
Plug sensor into station
Check wires in connector are not pulled out
There is an active leak/spill in
the bottom of the cart
Clean up spill/repair leak
Summary of Contents for Maestro MB1000 Series
Page 52: ...Installation Service User Manual Maestro series MB1000 MAN rev 1 52 9 2 Electrical diagrams...
Page 53: ...Installation Service User Manual Maestro series MB1000 MAN rev 1 53...
Page 54: ...Installation Service User Manual Maestro series MB1000 MAN rev 1 54...
Page 55: ...Installation Service User Manual Maestro series MB1000 MAN rev 1 55 9 3 Plumbing diagrams...
Page 56: ...Installation Service User Manual Maestro series MB1000 MAN rev 1 56...