Calibration using FDT/DTM Software
This section provides information on calibrating the STZ using a PC with PACTware (an FDT
frame application). For users who already have an FDT Frame Application, all the following
information is still relevant. For more information on FDT/DTM please refer to www.fdtgroup.org.
To perform the Input Trimming of the STZ set up equipment in Table 2.2 as shown in Figure 2.7.
Table 2.2.
Necessary Equipment
Figure 2.7.
Calibration Setup for PACTware
Accurate to ±0.05% of span for the intended application.
24Vdc, ±10%
250 ohms with ±0.01% precision
Calibrated to an accuracy of ±0.025%, or better; such as Fluke Model 725, or similar
Device
Specifications
Millivolt or
Resistance Source
Power Supply
Load Resistor
Multimeter
PACTware version 4.1 SP3 or newer
PC with PACTware
or FDT frame application
Moore Industries P/N 803-040-26 or 804-030-26
Communications Cable
For instructions on installation and setup of the STZ DTM with PACTware or another FDT frame,
please refer to section 5. See Figure 2.7 for the power supply and loop connections. The STZ
[HP] is shown, but the power supply and loop connections for the STZ [HPP] and STZ [DIN] will
be similar.
ADDR
ADDR
AD
STZ
12-24Vdc
Power
Supply
Load= 250 ohm
+
-
Moore Industries
P/N: 804-030-26
To USB
Port of PC
PC
HART Modem
connects to
Serial/USB
COM Port of PC
Functional Safety Dual Input Smart HART® Temperature Transmitter
STZ
SECTION 2