moonO LC-BP01 User Manual Download Page 5

CAUTION

 

 

 

 

LCD Display Signal

* The service life of the cuff may vary by the frequency of washing, skin condition, and 

storage state. The typical service life is 10000 times.

* It is recommended that the performance should be checked every 2 years and after 

maintenance and repair, by retesting at least the requirements in limits of the error of the cuff 

pressure indication and air leakage (testing at least at 50mmHg and 200mmHg).

* Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT according 

to the local guidelines.

* Manufacturer will make available on request circuit diagrams, component part lists, 

descriptions, calibration instructions,etc., to assist to service personnel in parts repair.

* The plug/adapter plug pins insulate the device from the main supply. Do not position the 

device in a position where it is difficult to disconnect from the supply mains to safely 
terminate operation of ME equipment.

* The operator shall not touch output of batteries /adapter and the patient simultaneously.

* Cleaning :Dust environment may affect the performance of the unit. Please use the soft 

cloth to clean the whole unit before and after use. Don’t use any abrasive or volatile cleaners.

* The device doesn’t need to be calibrated within two years of reliable service.

* If you have any problems with this device, such as setting up, maintaining or using, please 

contact the SERVICE PERSONNEL of moon. Don’t open or repair the device by yourself in 
the event of malfunctions. The device must only be serviced, repaired and opened by 
individuals at authorized sales/service centers.

* Please report to moon if any unexpected operation or events occur.

* Keep the unit out of reach of infant

s, young children or pets to avoid inhalation or 

swallowing of small parts. It is dangerous or even fatal.

* Be careful to strangulation due to cables and hoses, particularly due to excessive length.

* At least 30 min required for ME equipment to warm from the minimum storage temperature 

between uses until it is ready for intended use. At least 30 min required for ME equipment to 
cool from the maximum storage temperature between uses until it is ready for intended use.

* This 

equipment needs to be installed and put into service in accordance with the 

information provided in the ACCOMPANYING DOCUMENTS;

* Wireless communications equipment such as wireless home network devices, mobile 

phones, cordless telephones and their base stations, walkie-talkies can affect this 
equipment and should be kept at least a distance d away from the equipment. The distance 
d is calculated by the MANUFACTURER from the 80MHz to 5.8 GHz column of Table 4 and 
Table 9 of IEC 60601-1-2:2014, as appropriate.

* Please use ACCESSORIES and detachable parts specified/ authorised by 

MANUFACTURE. Otherwise, it may cause damage to the unit or danger to the user/patients.

* There is no luer lock connectors are used in the construction of tubing, there is a possibility 

that they might be inadvertently connected to intravascular fluid systems, allowing air to be 
pumped into a blood vessel.

* Please use the device under the environment which is provided in the user manual. 

Otherwise, the performance and lifetime of the device will be impacted and reduced.

INTRODUCTION

INTRODUCTION

6

7

SYMBOL

DESCRIPTION

EXPLANATION

Systolic blood pressure

The high pressure measured.

Diastolic blood pressure

The low pressure measured. 

User ID

Ba�ery Indicator

Indicate the current ba�ery. 

The bluetooth transfer icon blinks when the bluetooth is working.

Bluetooth transfer icon

appears when the monitor is operated by User 1.
appears when the monitor is operated by User 2.
appears when the monitor is operated by guest .

NOTE:  User 1 and User 2, each with 99 memory spaces;

User G, no momery space.

mmHg

Measurement Unit of the blood pressure.

Pulse/minute

Measurement Unit of

 Heart Rate.

Bluetooth transfer icon The bluetooth transfer icon blinks when the bluetooth is not working.

Voice broadcast

The function of voice broadcast is turned off

Voice broadcast

The function of voice broadcast is turned on

kpa

kPa

Measurement Unit of the blood pressure.(1kPa=7.5mmHg)

(1mmHg=0.133kPa)

Summary of Contents for LC-BP01

Page 1: ...usage of this product Keep the manual well for your further reference in case you have problems Thank you very much for selecting moon Blood Pressure Monitor LC BP01 Arm Type Version 1 0 Guangdong Tr...

Page 2: ...BLESHOOTING 20 SPECIFICATIONS 21 ATHORIZED COMPONENT 22 CONTACT INFORMATION 22 Safety Information Power Supply and Charge Power Setting the Time Date and Unit MEASUREMENT Start Measurement 14 COMPLIED...

Page 3: ...door use only F1 Thank you for selecting moon arm type blood pressure monitor LC BP01 The monitor features blood pressure measurement pulse rate measurement and the result storage The design provides...

Page 4: ...ent please avoid compression or restriction of the connection tubing The device cannot be used with HF surgical equipment at the same time CAUTION The ACCOMPANYING DOCUMENT shall disclose that the SPH...

Page 5: ...erature between uses until it is ready for intended use At least 30 min required for ME equipment to cool from the maximum storage temperature between uses until it is ready for intended use This equi...

Page 6: ...monitor at the cool dry and ventilated environment Avoid to approach to the fire and the heat source or it will cause the battery explode Only can use the moon s authorized AC Adaptor Model BLJ06L050...

Page 7: ...attempt to disassemble the blood pressure monitor or force open the built in battery Do not clean the blood pressure monitor when it is being charged Always unplug the charger first before cleaning t...

Page 8: ...able to insert one finger between the cuff and your arm Remove all jewelry such as watches and bracelets from your left arm Note If your doctor has diagnosed you with poor circulation in your left arm...

Page 9: ...measurement please relax first and then complete the measurement process Boot animation Before your start Inflating and measuring Display and save the results When the volume is on It will broadcast y...

Page 10: ...obtain the best performance please follow instructions below Measurements may be inaccurate if taken in the following circumstances Put in a dry place and avoid the sunshine Avoid intense shaking and...

Page 11: ...intervals the difference between each interval and the average is more than the average value of 25 or there are four or more pulse intervals the difference between each interval and the average is m...

Page 12: ...r no circumstance should you disassemble or attempt to repair the unit by yourself TROUBLESHOOTING SPECIFICATIONS Relax for a moment and then measure again The monitor detected motion talking or the p...

Page 13: ...mance Collateral standard Electromagnetic disturbances Requirements and tests EN ISO 81060 1 2012 Non invasive sphygmomanometers Part 1 Requirements and test methods for non automated measurement type...

Page 14: ...emissions RF emissions CISPR 11 RF emissions CISPR 11 Group 1 Class B Class A Comply Compliance Harmonic emissions IEC 61000 3 2 Voltage fluctuations flicker emissions IEC 61000 3 3 Emissions test Ta...

Page 15: ...n the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiv...