EMC Guidance

The ME EQUIPMENT or ME SYSTEM is suitable for home healthcare environments.

Warning: Don’t be near the active HF surgical equipment and the RF shielded room of an ME 
system for magnetic resonance imaging, where the intensity of EM disturbances is high.

Warning: Use of this equipment adjacent to or stacked with other equipment should be avoided 
because it could result in improper operation. If such use is necessary, this equipment and the 
other equipment should be observed to verify that they are operating normally.

Warning: Use of accessories, transducers and cables other than those specified or provided by 
the manufacturer of this equipment could result in increased electromagnetic emissions or 
decreased electromagnetic immunity of this equipment and result in improper operation.

Warning: Portable RF communications equipment (including peripherals such as antenna 

cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of 

the equipment LC-BP01, including cables specified by the manufacturer. Otherwise, 
degradation of the performance of this equipment could result.

Technical description

1. All necessary instructions for maintaining BASIC SAFETY and ESSENTIAL 
PERFORMANCE with regard to electromagnetic disturbances for the excepted service life.

2. Guidance and manufacturer’s declaration-electromagnetic emissions and Immunity.

Table 1 

Guidance and manufacturer’s declaration - electromagnetic emissions

RF emissions
CISPR 11 

RF emissions
CISPR 11 

Group 1 

Class [ B ]

Class A

Comply

Compliance

Harmonic emissions
IEC 61000-3-2 

Voltage fluctuations / 
flicker emissions
IEC 61000-3-3

Emissions test

Table 2 

 

Guidance and manufacturer’s declaration – electromagnetic Immunity 

Immunity Test

 

±8 kV contact
±2 kV, ±4 kV, ±8 kV,  ±15 kV air 

±8 kV contact
±2 kV, ±4 kV, ±8 kV,  ±15 kV air 

 

±2 kV for power supply lines
±1 kV signal input/output 
100 kHz repetition frequency

±2 kV for power supply lines
±1 kV signal input/output 
100 kHz repetition frequency

±0.5 kV, ±1 kV differential mode
±0.5 kV, ±1 kV, ±2 kV common mode    

±0.5 kV, ±1 kV differential mode
±0.5 kV, ±1 kV, ±2 kV common mode    

30 A/m
50 Hz / 60 Hz

30 A/m
50 Hz / 60 Hz

NOTE  U

is the a.c. mains voltage prior to application of the test level. 

Compliance level

Electrostatic

discharge (ESD) 

IEC 61000-4-2  

Power frequency 
magnetic field
IEC 61000-4-8

Voltage dips, short 
interruptions and 
voltage variations 
on power supply
 input lines 
IEC 61000-4-11

Electrical fast
transient/burst
IEC 61000-4-4   

Surge
IEC61000-4-5

IEC 60601-1-2 
Test level

0% U

T

; 0,5 cycle. At 0°, 45°, 90°, 135°,

180°, 225°, 270° and 315°.
0% U

T

; 1 cycle and 70% U

T

25/30 cycles; Single phase: at 0°.
0% U

T

; 250 / 300 cycle 

0% U

T

; 0,5 cycle. At 0°, 45°, 90°, 135°,

180°, 225°, 270° and 315°.
0% U

T

; 1 cycle and 70% U

T

25/30 cycles; Single phase: at 0°.
0% U

T

; 250 / 300 cycle 

Conduced RF
IEC61000-4-6

3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands 
between 0,15 MHz and 80 MHz
80% AM at 1 kHz

3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands 
between 0,15 MHz and 80 MHz
80% AM at 1 kHz

10 V/m
80 MHz – 2,7 GHz
80% AM at 1 kHz

10 V/m
80 MHz – 2,7 GHz
80% AM at 1 kHz

Radiated RF
IEC61000-4-3

EMC GUIDANCE

EMC GUIDANCE

Summary of Contents for LC-BP01

Page 1: ...usage of this product Keep the manual well for your further reference in case you have problems Thank you very much for selecting moon Blood Pressure Monitor LC BP01 Arm Type Version 1 0 Guangdong Tr...

Page 2: ...BLESHOOTING 20 SPECIFICATIONS 21 ATHORIZED COMPONENT 22 CONTACT INFORMATION 22 Safety Information Power Supply and Charge Power Setting the Time Date and Unit MEASUREMENT Start Measurement 14 COMPLIED...

Page 3: ...door use only F1 Thank you for selecting moon arm type blood pressure monitor LC BP01 The monitor features blood pressure measurement pulse rate measurement and the result storage The design provides...

Page 4: ...ent please avoid compression or restriction of the connection tubing The device cannot be used with HF surgical equipment at the same time CAUTION The ACCOMPANYING DOCUMENT shall disclose that the SPH...

Page 5: ...erature between uses until it is ready for intended use At least 30 min required for ME equipment to cool from the maximum storage temperature between uses until it is ready for intended use This equi...

Page 6: ...monitor at the cool dry and ventilated environment Avoid to approach to the fire and the heat source or it will cause the battery explode Only can use the moon s authorized AC Adaptor Model BLJ06L050...

Page 7: ...attempt to disassemble the blood pressure monitor or force open the built in battery Do not clean the blood pressure monitor when it is being charged Always unplug the charger first before cleaning t...

Page 8: ...able to insert one finger between the cuff and your arm Remove all jewelry such as watches and bracelets from your left arm Note If your doctor has diagnosed you with poor circulation in your left arm...

Page 9: ...measurement please relax first and then complete the measurement process Boot animation Before your start Inflating and measuring Display and save the results When the volume is on It will broadcast y...

Page 10: ...obtain the best performance please follow instructions below Measurements may be inaccurate if taken in the following circumstances Put in a dry place and avoid the sunshine Avoid intense shaking and...

Page 11: ...intervals the difference between each interval and the average is more than the average value of 25 or there are four or more pulse intervals the difference between each interval and the average is m...

Page 12: ...r no circumstance should you disassemble or attempt to repair the unit by yourself TROUBLESHOOTING SPECIFICATIONS Relax for a moment and then measure again The monitor detected motion talking or the p...

Page 13: ...mance Collateral standard Electromagnetic disturbances Requirements and tests EN ISO 81060 1 2012 Non invasive sphygmomanometers Part 1 Requirements and test methods for non automated measurement type...

Page 14: ...emissions RF emissions CISPR 11 RF emissions CISPR 11 Group 1 Class B Class A Comply Compliance Harmonic emissions IEC 61000 3 2 Voltage fluctuations flicker emissions IEC 61000 3 3 Emissions test Ta...

Page 15: ...n the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiv...

Reviews: