moonO LC-BP01 User Manual Download Page 12

21

20

3.7 V 1000 mAH Built-in rechargeable 
li-polymer battery, 5V       1A AC Adapter

Digital LCD display
52.00mm diameter circular 

Approx. 265 g

Approx. 214 mm × 74.5 mm × 42 mm

A01 

Power supply

Display mode

Measurement mode

Oscillographic testing mode

Measurement range

Accuracy

Normal working condition

Storage & transportation

condition

Measurement perimeter

of the arm

Weight

External dimensions

Attachment

Mode of operation

Continuous operation

Degree of protection

Type BF applied part

Protection against 

ingress of water

Software version

Device classification

Battery Powered Mode:
Internally Powered ME Equipment
AC Adaptor charged Mode: Class II ME Equipment

WARNING: No modification of this equipment is allowed.

Pressure:

5°C-40°C within ±3 mmHg (0.4 kPa) 

Pulse value: ±5%

Rated cuff pressure: 

0 mmHg ~ 299 mmHg (0 kPa ~ 39.9 kPa)

Measurement pressure: 

SYS: 60 mmHg ~ 230 mmHg (8.0 kPa~ 30.7 kPa) 

DIA: 40 mmHg ~ 130 mmHg (5.3 kPa ~ 17.3 kPa)

Pulse value: (40-199)beat/minute

About 22 cm ~ 32 cm or 22 cm ~ 42 cm

A temperature range of : +5°C to +40°C
A relative humidity range of 15% to 90%,
non-condensing, but not requiring a water
vapour partial pressure greater than 50 hPa
An atmospheric pressure range of :
700 hPa to 1060 hPa

Temperature: -20°C to +60°C

A relative humidity range of ≤ 93%, non-condensing,

at a water vapour pressure up to 50 hPa

If any abnormality arises during use, please check the following points:

NOTE: If the product still does not work, contact moon Customer Service. Under no 
circumstance should you disassemble or attempt to repair the unit by yourself. 

TROUBLESHOOTING

SPECIFICATIONS

Relax for a moment
and then measure again.

The monitor detected 
motion,talking or the 
pulse is too poor while 
measuring.

Out of measurement
range

Relax for a moment.
Refasten the cuff and then
measure again. If the
problem persists, contact
your physician.

PROBLEM SYMPTOM

CHECK THIS

REMEDY

No power

Error
message

Display will not
light up.

Power is exhausted.

Charge the power

Power is low.

Warning
message

Click the ON/OFF button,
please wait for 4 seconds
 to turn on the device

Don't press repeatedly

Unable to 
charge

Poor contact of charging
 port or the adapter is 
not right

Please plug in the adapter
 again and use 5V to
 output the power adapter.

"Low power, 
please charge 
in time" is 
displayed. 

Low power

Charge the power

"Fault Code E01" 
is displayed .

The pressure of the
cuff is excess or
air connector plus gets 
loose. .

Check theair connector 
plus ,

refasten the cuff .

"Fault Code E02" 
is displayed .

"Fault Code E03" 
is displayed .

The measurement 
process does not detect 

the pulse signal.

Loosen the clothing on 
the arm and  then 
measure again.

"Fault Code E04" 
is displayed .

A calibration error
occurred or out of 
measurement range or
parameter error and so 
on. 

Retake the measurement.
If theproblem persists, 
contact the retailer or our
customer service for 
further assistance.
Refer to the warranty for 
contact information and 
return instructions.

"Fault Code OUT" 
is displayed .

User manual, USB Cable,AC Adaptor  

IP21 It means the device could protected against 
solid foreign objects of 12.5mm and greater, and 
protect against vertically falling water drops.

Summary of Contents for LC-BP01

Page 1: ...usage of this product Keep the manual well for your further reference in case you have problems Thank you very much for selecting moon Blood Pressure Monitor LC BP01 Arm Type Version 1 0 Guangdong Tr...

Page 2: ...BLESHOOTING 20 SPECIFICATIONS 21 ATHORIZED COMPONENT 22 CONTACT INFORMATION 22 Safety Information Power Supply and Charge Power Setting the Time Date and Unit MEASUREMENT Start Measurement 14 COMPLIED...

Page 3: ...door use only F1 Thank you for selecting moon arm type blood pressure monitor LC BP01 The monitor features blood pressure measurement pulse rate measurement and the result storage The design provides...

Page 4: ...ent please avoid compression or restriction of the connection tubing The device cannot be used with HF surgical equipment at the same time CAUTION The ACCOMPANYING DOCUMENT shall disclose that the SPH...

Page 5: ...erature between uses until it is ready for intended use At least 30 min required for ME equipment to cool from the maximum storage temperature between uses until it is ready for intended use This equi...

Page 6: ...monitor at the cool dry and ventilated environment Avoid to approach to the fire and the heat source or it will cause the battery explode Only can use the moon s authorized AC Adaptor Model BLJ06L050...

Page 7: ...attempt to disassemble the blood pressure monitor or force open the built in battery Do not clean the blood pressure monitor when it is being charged Always unplug the charger first before cleaning t...

Page 8: ...able to insert one finger between the cuff and your arm Remove all jewelry such as watches and bracelets from your left arm Note If your doctor has diagnosed you with poor circulation in your left arm...

Page 9: ...measurement please relax first and then complete the measurement process Boot animation Before your start Inflating and measuring Display and save the results When the volume is on It will broadcast y...

Page 10: ...obtain the best performance please follow instructions below Measurements may be inaccurate if taken in the following circumstances Put in a dry place and avoid the sunshine Avoid intense shaking and...

Page 11: ...intervals the difference between each interval and the average is more than the average value of 25 or there are four or more pulse intervals the difference between each interval and the average is m...

Page 12: ...r no circumstance should you disassemble or attempt to repair the unit by yourself TROUBLESHOOTING SPECIFICATIONS Relax for a moment and then measure again The monitor detected motion talking or the p...

Page 13: ...mance Collateral standard Electromagnetic disturbances Requirements and tests EN ISO 81060 1 2012 Non invasive sphygmomanometers Part 1 Requirements and test methods for non automated measurement type...

Page 14: ...emissions RF emissions CISPR 11 RF emissions CISPR 11 Group 1 Class B Class A Comply Compliance Harmonic emissions IEC 61000 3 2 Voltage fluctuations flicker emissions IEC 61000 3 3 Emissions test Ta...

Page 15: ...n the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiv...

Reviews: