LCD DISPLAY/RESOLUTION/POSSIBLE CAUSE
S
INFUSION
COMPLETE
AIR-IN-LINE
ALARM DOWN
OCCLUSION
ALARM UP
OCCLUSION
ALARM HIGH UP
PRESSURE
ALARM SET NOT
INSTALLED
ALARM
UNATTENDED
ALARM REPLACE
SET 3
ALARM REPLACE
SET 4
ALARM DOOR
OPEN
ALARM EMPTY
BATTERY
ERROR CODE
Press PAUSE, attach new solution container, highlight REPEAT RX and press YES if
instructed to do so by your clinician. The program will auto-scroll (or confirm each
field by pressing YES). If therapy is complete, turn pump off.
Press the PRIME key, then follow on-screen instructions to remove air from the
tubing and IV bag if instructed to do so by your clinician. CAUTION: Disconnect
set from your access site before priming. Sets with in-line ASVs must be primed
on the pump.
Down occlusions auto-correct when occlusion is resolved without the need to
press PAUSE. Check the tubing from pump to your IV site for cause, then correct
obstruction if instructed to do so by your clinician.
Check tubing from pump to IV bag for cause, then correct obstruction if
instructed to do so by your clinician. Possible causes: glass container without
vented spike adapter, line kinked in pump door, bag empty, bag not spiked
properly.
Check for excessive pressure on IV bag. Causes: uneven fluid distribution within
lockbox or carrying pouch, object pressing on solution container.
Use only Curlin administration sets. Check that set is installed properly. Refer to
Instructions for Use in set package or Installing Administration Set section in this
instruction booklet. Possible causes: tab not removed from yellow Flow-Stop,
blue guide pin or yellow Flow-Stop not properly seated in receptacles, no set
installed.
The pump will alarm when left idle for 2 minutes. Press RUN to start the infusion
or press OFF to power down the pump.
Replace set and prime if necessary. Possible causes: flattening of the soft
pumping segment of the set due to overuse, use of a glass container without a
vented spike adapter, excessive plunger stiction with syringe use.
Equalize the pressure in the set by priming if instructed to do so by your clinician
(disconnect from your IV site). Turn pump OFF then ON to reset the sensor prior
to resuming the infusion.
Check placement of the administration set and close pump door properly.
Turn pump off by pressing the OFF key. Install 2 new batteries and turn pump
back on.
If an error code occurs, turn pump off and consult your clinician, or call the toll
free number on the back of the pump if instructed to do so by your clinician.
TROUBLESHOOTING:
NOTE: For all alarms except “INFUSION COMPLETE”, follow these instructions before proceeding
to the RESOLUTION section below: To silence and exit Alarm/Alert screens, press the PAUSE key,
press YES to RESUME, then resolve the alarm. When ready to resume the infusion, press RUN to
start the infusion.