6

 

STARTING A NEW MEDICATION CONTAINER (BAG)

1.  If the pump is turned off, turn the pump on by pressing the ON key.  Wait for the pump to go 

through a self-test.  This should only take about 20 seconds.   

2.   Press  YES to PROGRAM.  PROGRAM should be highlighted on the screen (or press YES to LIBRARY if 

instructed by your clinician).

3.   Press  YES to REPEAT RX if you have not started infusing from the current medication container.  

REPEAT RX should be highlighted on the screen.  If it is not highlighted, use the DOWN ARROW key 

to highlight REPEAT RX then press the YES key.

 

ALWAYS press YES to REPEAT RX when you are going to start a NEW medication bag that has never 

started infusing.  When you select REPEAT RX, the pump may ask questions to confirm and caution 

that you are resetting the pump for a new bag.  Press the YES key to all on-screen questions.  The 

pump will then automatically scroll through the program for you to review.  Allow the screen to scroll.  

If you press any key during the scrolling cycle, the screen will stop.  If this occurs, press the YES key to 

every field until the following screen is reached.   

Summary of Contents for curlin 6000 series

Page 1: ...MOOG Medical Devices Group Curlin 6000 Series Ambulatory Infusion Pump PATIENT GUIDE...

Page 2: ...oving the Tubing from the Pump 4 Priming the Tubing by Gravity 4 Loading the Tubing into the Pump 5 Starting a New Medication Container Bag 6 Priming the Administration Set Using the Pump 7 Restarting...

Page 3: ...u how to use them 1 ON OFF Used to turn the pump on and off 2 RUN PAUSE Used to start or pause the infusion When you are at the Run to Start screen this key will start the infusion This key will also...

Page 4: ...fy that the metal on the bottom of the batteries is in contact with the springs in the bottom of the battery compartment 9 If a plastic pull tab is present ensure that the pull tab is between the batt...

Page 5: ...k Slide Clamp Integral Flow Stop with breakaway tab Tubing Guide Pin Bag Spike These drawings are NOT to scale REMOVING THE ADMINISTRATION SET TUBING FROM THE PUMP 1 Open the pump door and remove the...

Page 6: ...oor of the pump lift the latch on the top of the door 2 Pull the door up and all the way to the right refer to the diagram on the next page 3 Locate the BLUE and YELLOW arrows within the pump as a ref...

Page 7: ...n the screen If it is not highlighted use the DOWN ARROW key to highlight REPEAT RX then press the YES key ALWAYS press YES to REPEAT RX when you are going to start a NEW medication bag that has never...

Page 8: ...ing Menu Screen 4 Press and hold the PRIME key You will hear the pump motor running and you will start to see the fluid moving through the tubing 5 If there is a filter on the tubing hold the filter u...

Page 9: ...to the pump align the blue tab on the power cord with the matching keyhole at the bottom of the pump The keyhole and receptacle are marked POWER in blue writing 3 Insert the port of the AC adapter int...

Page 10: ...again to exit the help screen Where does the remote Bolus Cable with a green ring plug into on the pump The port on the side of the pump labeled BOLUS DATA is for the remote Bolus Cable with switch T...

Page 11: ...n solution container Use only Curlin administration sets Check that set is installed properly Refer to Instructions for Use in set package or Installing Administration Set section in this instruction...

Page 12: ...l the Clinical Support line if instructed to do so by your clinician Clinical Support 1 800 970 2337 www moog com medical Moog is a registered trademark of Moog Inc All trademarks indicated herein are...

Reviews: