background image

ALARMS

When an alarm sounds you will hear a beep(s), the red alarm light will flash and the data screen will 

display the reason for the alarm. To resolve alarm situations refer to the Troubleshooting section.  

Pressing the SILENCE key will silence the alarm for one minute.  Pressing the PAUSE key will also 

silence the alarm.

 

FREQUENTLY ASKED QUESTIONS

What should I do if I have questions about the pump? Whenever you have questions about the pump 

or your therapy, you should call your healthcare provider.
How do I move around with the pump? Curlin Medical Pumps have a carry pack specifically designed 

to allow you to be mobile and active while receiving your therapy. Your healthcare provider will 

demonstrate how to use the carry pack.
Where do I leave the pump when I want to sleep? Place the pump at your bedside or next to you on 

the bed. Your clinician may have special instructions for bedtime use of the pump.
How do I bathe or shower during my therapy? The pump is water resistant, but not water proof, so it 

should be placed outside your tub or shower.
How do I exit the help screen? When the pump is paused and the HELP/OPTION key is pressed, a 

help message will appear. Press the HELP/OPTIONS key again to exit the help screen.
Where does the remote “Bolus Cable” (with a green ring) plug into on the pump? The port on the side 

of the pump (labeled BOLUS/DATA) is for the remote “Bolus Cable” with switch. The bolus cable is 

inserted and removed the same way as the AC adaptor.
Approximately how long will new batteries last? Battery life depends on the usage. The following 

times are approximate only:

• Nominal 85 hours at a rate of  2 ml/hr

• Nominal 30 hours at a rate of 125 ml/hr

• Nominal 10 hours at a rate of 400 ml/hr
However, pressing the keys frequently will illuminate the display and the battery life will be 

shortened. To extend battery life, the AC Adapter may be used at night.
How can I determine how much battery power is remaining in the pump while the pump is running? 

If you have the words “Power Check” listed next to the NUMBER 2 KEY, press it while the pump is 

running and the display will show you a bar graph indicating the battery power. This will remain 

illuminated for several seconds and will not affect the function of the pump. Otherwise, press the 

HELP/OPTIONS key and arrow down to “PWR CK” and press the NO key to change the field from 

OFF to ON. Then press the YES key to enter the menu which will show you the power bar graph. After 

several seconds the menu will revert to the Options screen. Press the YES key to accept options. 

Are the “ C” cell batteries in the pump rechargeable? No.

9

Summary of Contents for curlin 6000 series

Page 1: ...MOOG Medical Devices Group Curlin 6000 Series Ambulatory Infusion Pump PATIENT GUIDE...

Page 2: ...oving the Tubing from the Pump 4 Priming the Tubing by Gravity 4 Loading the Tubing into the Pump 5 Starting a New Medication Container Bag 6 Priming the Administration Set Using the Pump 7 Restarting...

Page 3: ...u how to use them 1 ON OFF Used to turn the pump on and off 2 RUN PAUSE Used to start or pause the infusion When you are at the Run to Start screen this key will start the infusion This key will also...

Page 4: ...fy that the metal on the bottom of the batteries is in contact with the springs in the bottom of the battery compartment 9 If a plastic pull tab is present ensure that the pull tab is between the batt...

Page 5: ...k Slide Clamp Integral Flow Stop with breakaway tab Tubing Guide Pin Bag Spike These drawings are NOT to scale REMOVING THE ADMINISTRATION SET TUBING FROM THE PUMP 1 Open the pump door and remove the...

Page 6: ...oor of the pump lift the latch on the top of the door 2 Pull the door up and all the way to the right refer to the diagram on the next page 3 Locate the BLUE and YELLOW arrows within the pump as a ref...

Page 7: ...n the screen If it is not highlighted use the DOWN ARROW key to highlight REPEAT RX then press the YES key ALWAYS press YES to REPEAT RX when you are going to start a NEW medication bag that has never...

Page 8: ...ing Menu Screen 4 Press and hold the PRIME key You will hear the pump motor running and you will start to see the fluid moving through the tubing 5 If there is a filter on the tubing hold the filter u...

Page 9: ...to the pump align the blue tab on the power cord with the matching keyhole at the bottom of the pump The keyhole and receptacle are marked POWER in blue writing 3 Insert the port of the AC adapter int...

Page 10: ...again to exit the help screen Where does the remote Bolus Cable with a green ring plug into on the pump The port on the side of the pump labeled BOLUS DATA is for the remote Bolus Cable with switch T...

Page 11: ...n solution container Use only Curlin administration sets Check that set is installed properly Refer to Instructions for Use in set package or Installing Administration Set section in this instruction...

Page 12: ...l the Clinical Support line if instructed to do so by your clinician Clinical Support 1 800 970 2337 www moog com medical Moog is a registered trademark of Moog Inc All trademarks indicated herein are...

Reviews: