IPA287Supplement Page
3
Monroe Electronics, Inc
.
---V A30%M50%
28C 9.0V 34%RH
Main entrance display
+Decay
28C 9.0V 34%RH
Balance
28C 9.0V 34%RH
Null
28C 9.0V 34%RH
Press [+] and hold
less than 2 seconds
between 2-4 seconds
longer than 4 seconds
+ 1V Zeroing
28C 9.0V 34% M1+
+ 950V +Charge
28C 9.0V 34% M1+
+ 1V Zeroing
28C 9.0V 34% M2B
+1030V +Decay
28C 9.0V 34% M1+
+ 800V 3.6s
28C 9.0V 34% M1+
Manual
Test #1 (1-0)
+Decay
Manual
Test #2 (1-0)
Balance
+ 1V 5.2s
+0002mx –0004mn
Voltage update Balance duration timer
Min and Max voltages
start recording after 5 seconds.
Press [Enter] will abort any on-going test.
After the test complete, review data begins.
Press [^] or [v] will scroll through the data.
Press [Enter] or [ON/OFF] back to
Main
Manual Plus Decay Test
(press [^] button less than 2 seconds)
The 287 unit will start by zeroing then Charge to over +1050V, then Decay.
The decay timer starts at +1000V and stops at +100V.
Manual Balance Test
(press and hold [^] or [v] button for 2-4 seconds)
The 287 will start by zeroing, then Balance for the time duration. Start recording Min and Max
voltage levels after 5 seconds.
Up to 10 Manual test results can be stored in memory. They are named M1 thru M0 where M0 is
the 10
th
test.
Timer starts at
1000V and stops
at 100V