background image

IPA287Supplement Page 

3

 

Monroe Electronics, Inc

 

 
 

 
 

---V   A30%M50%

28C 9.0V 34%RH  

Main entrance display

+Decay

 

28C 9.0V 34%RH  

 

Balance
28C 9.0V 34%RH  

Null

 

28C 9.0V 34%RH  

Press [+] and hold

less than 2 seconds

 

between 2-4 seconds

longer than 4 seconds

+  1V   Zeroing

 

28C 9.0V 34% M1+  

 

+ 950V   +Charge

 

28C 9.0V 34% M1+  

 

+   1V   Zeroing

28C 9.0V 34% M2B  

+1030V   +Decay

 

28C 9.0V 34% M1+  

 

+ 800V     3.6s

 

28C 9.0V 34% M1+  

 

Manual

 

Test #1 (1-0)
+Decay

 

Manual

 

Test #2 (1-0)

 

Balance

 

+   1V    5.2s
+0002mx –0004mn  

Voltage update   Balance duration timer
Min and Max voltages

 

start recording after 5 seconds.

Press [Enter] will abort any on-going test.

 

After the test complete, review data begins.

 

Press [^] or [v] will scroll through the data.

 

Press [Enter] or [ON/OFF] back to 

Main

 

Manual Plus Decay Test 

 

(press [^] button less than 2 seconds)

The 287 unit will start by zeroing then Charge to over +1050V, then Decay.
The decay timer starts at +1000V and stops at +100V. 

Manual Balance Test 

 

(press and hold [^] or [v] button for 2-4 seconds)

The 287 will start by zeroing, then Balance for the time duration. Start recording Min and Max  

 

voltage levels after 5 seconds.

 

Up to 10 Manual test results can be stored in memory.  They are named M1 thru M0 where M0 is 
the 10

th

 test. 

Timer starts at 
1000V and stops
at 100V 