IPA287Supplement Page
5
Monroe Electronics, Inc
.
Auto Sequence Group Test
(press [AUTO] button less than 2 seconds)
The 287 unit will perform up to 10 groups of automatic test sequences and store the results in memory.
Groups are named 1 thru 0 where 0 is the 10
th
. Each group has 3 +decay, 3-decay, and an optional balance
test. The sequence can be set in the Set-Up Menu.
Decay test
: Zeroing to <5V, Charge to over +/-1050V, then Decay.
The decay timer starts at +/-1000V and stops at +/-100V.
Balance Test:
Zeroing to <5V, then Balance for the time duration. Start recording
Voltage peaks after 5 seconds.
The delay time between the tests can be set in the Set-Up Menu.
---V A30%M50%
28C 9.0V 34%RH
Main entrance display
AUTO
28C 9.0V 34%RH
Null
28C 9.0V 34%RH
Press [AUTO] and hold
less than 2 seconds
longer than 2 seconds
+ 1V Zeroing
28C 9.0V 34% 11+
+ 950V +Charge
28C 9.0V 34% 11+
Timer starts at 1000V
and stops at 100V.
Group #1 (1-0)
Test #1 (1-3)
+Decay
+ 1V Zeroing
28C 9.0V 34% 1Ba
Group #1 (1-0)
Balance test
+ 1V 5.2s
+0002mx –0004mn
Voltage update Balance duration timer
Max and min voltages start recording
after 5 seconds.
Press [Enter] will abort any on going test and back to
Main
.
After the test complete, review data begins.
Press [^] or [v] will scroll through the data.
Press [Enter] or [ON/OFF] back to
Main
Auto Sequence Test
go on for the other 5 decay tests...
Review test results
+1030V +Decay
28C 9.0V 34% 11+
+ 800V 3.6s
28C 9.0V 34% 11+
+ 20V 8.6s
Delay 6 sec