Monroe Electronics 287a Supplementary Manual Download Page 6

IPA287Supplement Page 

5

 

Monroe Electronics, Inc

 

 
 
 
 
 

Auto Sequence Group Test 

(press [AUTO] button less than 2 seconds)

 

The 287 unit will perform up to 10 groups of automatic test sequences and store the results in memory.  
Groups are named 1 thru 0 where 0 is the 10

th

.  Each group has 3 +decay, 3-decay, and an optional balance 

test. The sequence can be set in the Set-Up Menu.

 

Decay test

 

: Zeroing to <5V, Charge to over +/-1050V, then Decay.

The decay timer starts at +/-1000V and stops at +/-100V. 

Balance Test:

 

 Zeroing to <5V, then Balance for the time duration. Start recording 

 

Voltage peaks after 5 seconds.

The delay time between the tests can be set in the Set-Up Menu.

---V   A30%M50%

28C 9.0V 34%RH  

Main entrance display

AUTO

 

28C 9.0V 34%RH  

 

Null

 

28C 9.0V 34%RH  

Press [AUTO] and hold

less than 2 seconds

 

longer than 2 seconds

+   1V   Zeroing

 

28C 9.0V 34% 11+  

 

+ 950V   +Charge

 

28C 9.0V 34% 11+  

 

Timer starts at 1000V
and stops at 100V.

Group #1 (1-0)
Test #1 (1-3)
+Decay

+   1V   Zeroing
28C 9.0V 34% 1Ba  

Group #1 (1-0)

 

Balance test

 

+   1V    5.2s
+0002mx –0004mn  

 Voltage update   Balance duration timer
  Max and min voltages start recording
  after 5 seconds.

Press [Enter] will abort any on going test and back to 

Main

.

After the test complete, review data begins.

 

Press [^] or [v] will scroll through the data.

 

Press [Enter] or [ON/OFF] back to 

Main

 

Auto Sequence Test

 

go on for the other 5 decay tests...

 

Review test results

+1030V   +Decay

 

28C 9.0V 34% 11+  

 

+ 800V     3.6s

 

28C 9.0V 34% 11+  

 

+  20V   8.6s

 

Delay   6 sec  

 

Reviews: