IPA287Supplement Page
4
Monroe Electronics, Inc
.
---V A30%M50%
28C 9.0V 34%RH
Main entrance display
-Decay
28C 9.0V 34%RH
Balance
28C 9.0V 34%RH
Null
28C 9.0V 34%RH
Press [-] and hold
less than 2 seconds
between 2-4 seconds
longer than 4 seconds
+ 1V Zeroing
28C 9.0V 34% M3-
- 950V -Charge
28C 9.0V 34% M3-
-1030V -Decay
28C 9.0V 34% M3-
- 800V 3.6s
28C 9.0V 34% M3-
+ 1V
Zeroing
28C 9.0V 34% M4B
Manual
Test #3 (1-0)
-Decay
Manual
Test #4 (1-0)
Balance
+1V 5.2s
+0002mx –0004mn
Voltage update Balance duration timer
min and max voltages start recording
after 5 seconds.
Press [Enter] will abort any on going test.
After the test complete, review data begins.
Press [^] or [v] will scroll through the data.
Press [Enter] or [ON/OFF] back to
Main
Manual Minus Decay Test
(press [v] button less than 2 seconds)
The 287 unit will start by zeroing then Charge to over +1050V, then Decay.
The decay timer starts at +1000V and stops at +100V.
Manual Balance Test
(press and hold [^] or [v] button for 2-4 seconds)
The 287 will start by zeroing, then Balance for the time duration. Start recording Min and Max
peaks after 5 seconds.
Timer starts at
1000V and stops
at 100V