background image

IPA287Supplement Page 

4

 

Monroe Electronics, Inc

 

 
 

 
 

---V   A30%M50%

28C 9.0V 34%RH  

Main entrance display

-Decay

 

28C 9.0V 34%RH  

 

Balance
28C 9.0V 34%RH  

Null

 

28C 9.0V 34%RH  

Press [-] and hold

less than 2 seconds

 

between 2-4 seconds

longer than 4 seconds

+   1V   Zeroing

 

28C 9.0V 34% M3-  

 

- 950V   -Charge

 

28C 9.0V 34% M3-  

 

-1030V   -Decay

 28C 9.0V 34% M3-  

 

- 800V     3.6s

28C 9.0V 34% M3-  

 

+   1V   

Zeroing

28C 9.0V 34% M4B  

Manual

 

Test #3 (1-0)
-Decay

 

Manual
Test #4 (1-0)

 

Balance

+1V  5.2s
+0002mx –0004mn  

Voltage update   Balance duration timer
min and max voltages start recording
after 5 seconds.

 

Press [Enter] will abort any on going test.

 

After the test complete, review data begins.

 

Press [^] or [v] will scroll through the data.

 

Press [Enter] or [ON/OFF] back to 

Main

 

Manual Minus Decay Test 

 

(press [v] button less than 2 seconds)

The 287 unit will start by zeroing then Charge to over +1050V, then Decay.
The decay timer starts at +1000V and stops at +100V.

Manual Balance Test 

 

(press and hold [^] or [v] button for 2-4 seconds)

The 287 will start by zeroing, then Balance for the time duration. Start recording Min and Max  

 

peaks after 5 seconds.

 

Timer starts at 
1000V and stops
at 100V 