Time Setting :
1. Once the user sets the mode (
M1
or
M3
), please press
TIME
button
continuously for longer than 2 seconds (till the number shown in the LCD
panel starts to blink), and press again to change time, one second per push.
2. (
M1
and
M3
) Operating time will start to add up after one second if
the user pushes the TIME button continuously. Press
ON/OFF/SET
button to complete time setting.
3. The time setting of
M2
is for 15 minutes which can not be re-set.
Operation :
1. After setting up the MODE selection and operating time,
triggers the
ON/OFF/SET
button to start curing cycle.
2. Press again
ON/OFF/SET
button to stop operation at any time.
3. Auto Power Saving Mode: the device will turn off automatically if it is left
idled for longer than 90 seconds. Operation resumes once the user presses
the trigger
ON/OFF/SET
.
4. Under “
M1, 8 seconds
” setting, the device can be continuously ultilized
for approximately 30 curing sessions.
Caution :
Operating GT-1500 at
Mode 1
, 8 seconds for more than 30 cycles
continuously could result in high temperature accumulation, which is
hazardous to the patient’s teeth. The user should pay more caution to
check if any over-heating or contraindication.
COOLING SYSTEM
Ti-Lite
GT-1500
is equipped with a special built-in magnetic fan
motor to prevent over-heating and to prolong the light emitting time.
Radiometer Sensor :
1. Choose
M1
(full power)
2. Instert the Fiber-Optic tip to the handpiece.
3. Aim the tip at the
Sensor of Radiometer
and
press
ON/OFF/SET
button to start checking the
light intensity.
Over-heating
Indication
CHARGER SIGNALS
1. When the power of battery is low, place the handpiece onto
the charger stand in up-right position to begin charging.
2. When the charger starts charging, the
red
(CHARGE)
light will be turned on to indicate the charging begins.
3. The
green
(READY)
light resumes after the
handpiece is fully charged.
IMPORTANT :
> The cooling system will be
activated with the motor fan
which starts turning when the
heat starts to accumulate.
> The thermometer symbol will
start to blink and shut off the
system for complete cooling
and for protection purpose.