2.5
MEASURE BOLTS
You are now ready to do measurements.
Attach the transducer to the bolt. Make sure to apply a thin
layer of ultrasonic couplant between the bolt and transducer
and carefully rock the transducer onto the center of the bolt
head or end. Then press
MEASURE.
Please note: There can be no air gap between the transducer
and the bolt. Ultrasound do not travel in air.
If the bolt surface has protruding marking, this must be
grinded down or otherwise removed. Please see YouTube
video on bolt ends and surface roughness.
If no temperature probe is attached, push the temperature
button and manually enter the temperature. Then push
CONTINUE.
Let the instrument work for a few seconds while it determines
the best echo peak for this bolt. The LED-light will flash
red/green while working.
Standard Magnetic Transducer