FilterMax F3 and FilterMax F5 Multi-Mode Microplate Readers User Guide
34
5016208 D
7. On the front panel of the instrument, press
to move the plate drawer outside of
the instrument.
8. Remove the plate from the drawer and leave the plate drawer open.
9. Power off the instrument.
10. Unplug the power cord from the rear of the instrument and from the wall outlet.
11. Disconnect the USB cable from the rear of the instrument and from the computer.
12. Align the holes in the plate drawer transport lock with the holes on the end of the plate
drawer.
Item
Description
1
Screw #1 fastens the lock to the internal frame of the instrument
2
Screw #2: Fastens the lock to the plate drawer
3
Screw #3: Fastens the lock to the plate drawer
4
Transport lock
5
Plate drawer
6
Plate chamber door open
13. Use the 2.0 mm hex key to tighten screws #2 and #3 on the transport lock to attach the
lock to the plate drawer.
14. Gently push the plate drawer into the instrument and as far to the left as possible until
screw #1 lines up with the hole on the internal frame. Hold the plate chamber door open
until you fasten the transport lock.
15. Tighten screw #1 to lock the plate drawer in place.